粒子治疗用碳离子束的布拉格曲线模拟:使用GEANT4工具包的研究

Bragg-Curve Simulation of Carbon-Ion Beams for Particle-therapy Applications: a study with the GEANT4 toolkit

M. Kh. Hamad

摘要:

我们使用GEANT4 Monte Carlo MC Toolkit模拟碳离子束入射到水、组织和骨骼上,同时考虑到核碎裂反应。在增加初级射束的能量时,布拉格峰的位置转移到体模内部更深的位置。对于不同的材料,峰值位于沿束方向的较浅的深度处,并且随着电子密度NZ的增加而变得尖锐。随后,产生的深度剂量的布拉格曲线,然后基准与德国GSI的实验数据。计算结果与GSI实验数据具有较好的相关性,精度在0.02~0.08cm之间,为在重离子治疗计划中采用MC奠定了基础。Kolmogorov-Smirnov K-S检验从统计学的角度进一步确定了模拟数据与实验测量数据非常匹配。将入口处的二维等剂量线与峰值位置周围和超出峰值的尾部区域中的等剂量线进行比较,表明由于二次剂量,与水和组织相比,骨产生更多剂量。在水中,结果表明,每个碎片沉积的最大能量主要归因于次级碳离子,其次是次级硼和铍。此外,产生的质子数是最高的,从而对尾区的总剂量沉积做出最大贡献。最后分析了中子和光子的关联谱。平均中子能量值为16.29 MeV,次级γ射线为1.03 MeV。然而,中子剂量被认为是可以忽略不计的总剂量相比,由于其较长的范围。

英文摘要:

We used the GEANT4 Monte Carlo MC Toolkit to simulate carbon ion beams incident on water, tissue, and bone, taking into account nuclear fragmentation reactions. Upon increasing the energy of the primary beam, the position of the Bragg-Peak transfers to a location deeper inside the phantom. For different materials, the peak is located at a shallower depth along the beam direction and becomes sharper with increasing electron density NZ. Subsequently, the generated depth dose of the Bragg curve is then benchmarked with experimental data from GSI in Germany. The results exhibit a reasonable correlation with GSI experimental data with an accuracy of between 0.02 and 0.08 cm, thus establishing the basis to adopt MC in heavy-ion treatment planning. The Kolmogorov-Smirnov K-S test further ascertained from a statistical point of view that the simulation data matched the experimentally measured data very well. The two-dimensional isodose contours at the entrance were compared to those around the peak position and in the tail region beyond the peak, showing that bone produces more dose, in comparison to both water and tissue, due to secondary doses. In the water, the results show that the maximum energy deposited per fragment is mainly attributed to secondary carbon ions, followed by secondary boron and beryllium. Furthermore, the number of protons produced is the highest, thus making the maximum contribution to the total dose deposition in the tail region. Finally, the associated spectra of neutrons and photons were analyzed. The mean neutron energy value was found to be 16.29 MeV, and 1.03 MeV for the secondary gamma. However, the neutron dose was found to be negligible as compared to the total dose due to their longer range.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑