治疗用碳离子辐射径迹结构的纳米剂量学研究。第二部分:详细的辐射输运与径迹结构模拟
Nanodosimetric investigation of the track structure of therapeutic carbon ion radiation. Part 2: Detailed radiation transport and track structure simulation
AI总结:
本研究通过Geant4和PTra模拟,揭示治疗碳离子纳米剂量测量偏差源于束流几何与触发探测器限制,并提出改进实验设计的建议。
摘要:
先前报道的关于治疗能量碳离子穿透模拟组织的纳米剂量学测量结果,与纳米剂量计中碳离子的预测平均能量以及先前使用较低能量单能束进行的实验不相符。本研究旨在探讨这些差异的根源。使用Geant4工具包进行了详细模拟,以研究纳米剂量计中的辐射场,并为径迹结构模拟提供输入数据,这些模拟使用开发版本的PTra代码进行。Geant4模拟表明,在实验中采用的窄束几何条件下,只有一小部分碳离子穿过纳米剂量计,且其平均能量比目标值低12%至30%。这些碳离子中只有约三分之一或更少击中触发探测器。径迹结构模拟表明,观察到的增强电离簇大小主要归因于碳离子未击中触发探测器的事件与这些事件的巧合。此外,对于高吸收体厚度下碳离子穿过靶体积时观察到的差异,可以通过假设厚度或相互作用截面增加约1%来解释。结果表明,即使辐射场经过强准直,未来对临床碳离子束的纳米剂量学测量也需要大型触发探测器,以记录所有碳离子穿过纳米剂量计的事件。主束在吸收体中的能量损失计算不足,在规划此类实验时应以详细模拟替代。模拟代码中相互作用截面的不确定性可能会移动布拉格峰的位置。
英文摘要:
Previously reported nanodosimetric measurements of therapeutic-energy carbon ions penetrating simulated tissue have produced results that are incompatible with the predicted mean energy of the carbon ions in the nanodosimeter and previous experiments with lower energy monoenergetic beams. The purpose of this study is to explore the origin of these discrepancies. Detailed simulations using the Geant4 toolkit were performed to investigate the radiation field in the nanodosimeter and provide input data for track structure simulations, which were performed with a developed version of the PTra code. The Geant4 simulations show that with the narrow-beam geometry employed in the experiment, only a small fraction of the carbon ions traverse the nanodosimeter and their mean energy is between 12 % and 30 % lower than the targeted values. Only about one-third or less of these carbon ions hit the trigger detector. The track structure simulations indicate that the observed enhanced ionization cluster sizes are mainly due to coincidences with events in which carbon ions miss the trigger detector. In addition, the discrepancies observed for high absorber thicknesses of carbon ions traversing the target volume could be explained by assuming an increase in thickness or interaction cross-sections in the order of 1 %. The results show that even with strong collimation of the radiation field, future nanodosimetric measurements of clinical carbon ion beams will require large trigger detectors to register all events with carbon ions traversing the nanodosimeter. Energy loss calculations of the primary beam in the absorbers are insufficient and should be replaced by detailed simulations when planning such experiments. Uncertainties of the interaction cross-sections in simulation codes may shift the Bragg peak position.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。