SOI图像传感器对治疗用碳离子束的响应

Response of SOI image sensor to therapeutic carbon ion beam

Akihiko Matsumura

摘要:

碳离子放射疗法被认为是一种侵入性较小的癌症治疗方法。辐射质量是从测量的物理剂量中评价生物效应和临床剂量的重要参数。SOPHIAS探测器是一种具有宽动态范围和大有效面积的SOI图像传感器,在群马大学重离子医学中心(GHMC)使用治疗碳离子束测试了其性能。结果表明,SOPHIAS探测器可以区分一次碳和二次粒子。另一方面,特别是在高LET区域观察到LET依赖性。将通过使用器件模拟器和蒙特卡罗模拟来研究这种现象。

英文摘要:

Carbon ion radiotherapy is known as a less invasive cancer treatment. The radiation quality is an important parameter to evaluate the biological effect and the clinical dose from the measured physical dose. The performance of SOPHIAS detector, which is the SOI image sensor having a wide dynamic range and large active area, was tested by using therapeutic carbon ion beam at Gunma University Heavy Ion Medical Center (GHMC). It was shown that the primary carbon and secondary particles can be distinguishable by SOPHIAS detector. On the other hand, a LET dependence was observed especially at the high LET region. This phenomenon will be studied by using the device simulator together with Monte Carlo simulation.