通过体模研究对原型全视野光子计数CT系统的综合评估

Comprehensive evaluations of a prototype full field-of-view photon counting CT system through phantom studies

Xiaohui Zhan, Ruoqiao Zhang, Xiaofeng Niu, Ilmar Hein, Brent Budden, Shuoxing Wu, Nicolay Markov, Cameron Clarke, Yi Qiang, Hiroki Taguchi, Keiichi Nomura, Yosh… 展开作者

Xiaohui Zhan, Ruoqiao Zhang, Xiaofeng Niu, Ilmar Hein, Brent Budden, Shuoxing Wu, Nicolay Markov, Cameron Clarke, Yi Qiang, Hiroki Taguchi, Keiichi Nomura, Yoshihisa Muramatsu, Zhou Yu, Tatsushi Kobayashi, Richard Thompson, Hiroaki Miyazaki, Hiroaki Nakai

AI总结:

本研究评估了首个基于CdZnTe的原型全视野光子计数CT系统,证明其相比传统EID-CT在噪声、空间分辨率、物质定量及对比噪声比方面均具显著优势,并能生成超高分辨率图像。

摘要:

在过去二十年中,光子计数CT(PCCT)一直是研究热点。最近的研究和进展表明,使用基于半导体的光子计数探测器(PCD)的系统,与传统的基于闪烁体的系统相比,具有提供更好对比度、噪声和空间分辨率性能的潜力。通过多能阈值检测,PCD可以同时提供光子能量测量,并实现用于光谱成像的物质分解。在本研究中,我们报告了通过各种体模成像研究对我们首个基于CdZnTe的原型全尺寸光子计数CT系统进行的性能评估。该原型系统支持500 mm扫描视野(FOV)和等中心点10 mm锥形覆盖范围。体模扫描使用120 kVp、50至400 mAs的条件进行采集,以评估以下方面的成像性能:CT值准确性、均匀性、噪声、空间分辨率、物质区分和定量。定性和定量评估均表明,与传统的能量积分探测器(EID)-CT相比,PCCT具有卓越的成像性能,噪声更低且空间分辨率更高。使用投影域物质分解方法和多能量箱测量,PCCT虚拟单能图像(VMI)具有更低的噪声,并在碘和钙浓度定量方面表现更优。这些改进使得与EID-CT相比,高对比度和低对比度研究对象的对比噪声比(CNR)均有所提高。PCCT还可以使用比EID-CT小得多的探测器像素尺寸生成超高分辨率(SHR)图像,并显著提高图像空间分辨率。

英文摘要:

Photon counting CT (PCCT) has been a research focus in the last two decades. Recent studies and advancements have demonstrated that systems using semiconductor-based photon counting detectors (PCDs) have the potential to provide better contrast, noise and spatial resolution performance compared to conventional scintillator-based systems. With multi-energy threshold detection, PCD can simultaneously provide the photon energy measurement and enable material decomposition for spectral imaging. In this work, we report a performance evaluation of our first CdZnTe-based prototype full-size photon counting CT system through various phantom imaging studies. This prototype system supports a 500 mm scan field-of-view (FOV) and 10 mm cone coverage at isocenter. Phantom scans were acquired using 120 kVp from 50 to 400 mAs to assess the imaging performance on: CT number accuracy, uniformity, noise, spatial resolution, material differentiation and quantification. Both qualitative and quantitative evaluations show that PCCT has superior imaging performance with lower noise and improved spatial resolution compared to conventional energy integrating detector (EID)-CT. Using projection domain material decomposition approach with multiple energy bin measurements, PCCT virtual monoenergetic images (VMIs) have lower noise, and superior performance in quantifying iodine and calcium concentrations. These improvements lead to increased contrast-to-noise ratio (CNR) for both high and low contrast study objects compared to EID-CT. PCCT can also generate super-high resolution (SHR) images using much smaller detector pixel size than EID-CT and dramatically improve image spatial resolution.