Multi-contrast K-edge imaging on a bench-top photon-counting CT system: Acquisition parameter study
摘要:
目的:光子计数计算机断层成像(PCCT)显示出在材料分离和多种造影剂成像方面的医学成像前景。然而,许多PCCT设置正在开发中,并且没有针对特定的造影剂或用例进行优化。在这里,我们演示了如何改变实验系统参数以增强性能,并提出了一组基于造影剂实现这一目标的建议。方法:一个桌面PCCT系统与碲锌镉(CZT)探测器能够分离六个能量箱被用来图像多种造影剂在一个小的幻影。分离造影剂,并使用K边缘减法定量浓度。为了提高系统性能,我们研究了三个参数:射束滤波器类型和厚度,投影采集时间和能量箱宽度。基于PCCT信号和对比噪声比(CNR)或K边缘图像中的噪声比较参数的结果。在K边缘图像中定量造影剂的浓度,并与已知浓度进行比较。结果:台式PCCT系统能够通过K边缘减影成功地定量对比剂。减少投影采集时间显示K边缘CNR降低。然而,它并不以时间的平方根为尺度。滤波器类型和箱宽表现出对特定造影剂的依赖性。结论:所提出的台式系统证明了使用K边缘减法分离造影剂并准确确定K边缘图像中的造影剂浓度的能力。将根据造影剂选择未来使用的具体参数。
英文摘要:
Purpose: Photon-counting computed tomography (PCCT) shows promise for medical imaging in regards to material separation and imaging of multiple contrast agents. However, many PCCT setups are under development and are not optimized for specific contrast agents or use cases. Here, we demonstrate how experimental system parameters may be varied in order to enhance performance and we propose a set of recommendations to achieve this based on contrast agent. Approach: A table-top PCCT system with a cadmium zinc telluride (CZT) detector capable of separating six energy bins was used to image multiple contrast agents in a small phantom. The contrast agents were separated and the concentration was quantified using K-edge subtraction. To increase system performance, we investigated three parameters: beam filter type and thickness, projection acquisition time, and energy bin width. The results from the parameters were compared based on PCCT signal and contrast to noise ratio (CNR) or noise in K-edge images. The concentrations of the contrast agents were quantified in K-edge images and compared to known concentrations. Results: The bench-top PCCT system was able to successfully quantify the contrast agents through K-edge subtraction. Decreasing projection acquisition time showed a decrease in K-edge CNR. However, it did not scale as the square root of time. Filter type and bin width demonstrated a dependence on the specific contrast agent. Conclusions: The presented bench-top system demonstrated the ability to separate contrast agents using K-edge subtraction and accurately determine contrast concentration in K-edge images. Specific parameters for future use will be chosen based on contrast agent.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。