arXiv 1803.06008physics.med-ph
收藏

用于材料成分的定量双能和三能CT成像的统一图像重建框架

A unified image reconstruction framework for quantitative dual- and triple-energy CT imaging of material compositions

Wei Zhao, Don Vernekohl, Fei Han, Bin Han, Hao Peng, Lei Xing, James K Min

摘要:

许多临床应用严重依赖于患者解剖结构中不同类型材料的准确区分和分类。这项工作介绍了一个统一的框架,准确的非线性材料分解,并将其应用于第一次,在三能CT(TECT)的概念,增强材料的分化和分类,以及双能CT。三能量数据采集是在微型CT和临床CT成像的规模与商业“TwinBeam”双源DECT配置和快速kV开关DECT配置。材料分解和定量比较与光子计数检测器和蝴蝶结过滤器的存在下也进行。所提出的方法提供了定量的材料和能量选择性的图像检查现实的配置双和三能CT测量。与多色kV CT图像相比,虚拟单色图像显示出更好的图像质量。对于小鼠体模,定量测量表明,使用TECT和理想化光子计数CT(PCCT)获得的钆双胺和碘浓度之间的差异分别小于8 mg/mL和1 mg/mL。TECT在多对比度CT成像方面优于DECT,并且在频谱估计方面具有鲁棒性。对于胸部体模,对于所有实际配置,对比度图的浓度与相应真实参考值之间的差异均小于7 mg/mL。一个统一的框架,双能和三能CT成像已经建立了准确提取的材料成分,考虑到目前可用的商业DECT配置。该新技术有望为几种基于CT的诊断和治疗应用提供迫切需要的解决方案。

英文摘要:

Many clinical applications depend critically on the accurate differentiation and classification of different types of materials in patient anatomy. This work introduces a unified framework for accurate nonlinear material decomposition and applies it, for the first time, in the concept of triple-energy CT (TECT) for enhanced material differentiation and classification as well as dual-energy CT. The triple-energy data acquisition is implemented at the scales of micro-CT and clinical CT imaging with commercial "TwinBeam" dual-source DECT configuration and a fast kV switching DECT configuration. Material decomposition and quantitative comparison with a photon counting detector and with the presence of a bow-tie filter are also performed. The proposed method provides quantitative material- and energy-selective images examining realistic configurations for both dual- and triple-energy CT measurements. Compared to the polychromatic kV CT images, virtual monochromatic images show superior image quality. For the mouse phantom, quantitative measurements show that the differences between gadodiamide and iodine concentrations obtained using TECT and idealized photon counting CT (PCCT) are smaller than 8 mg/mL and 1 mg/mL, respectively. TECT outperforms DECT for multi-contrast CT imaging and is robust with respect to spectrum estimation. For the thorax phantom, the differences between the concentrations of the contrast map and the corresponding true reference values are smaller than 7 mg/mL for all of the realistic configurations. A unified framework for both dual- and triple-energy CT imaging has been established for the accurate extraction of material compositions; considering currently available commercial DECT configurations. The novel technique is promising to provide an urgently needed solution for several CT-based diagnosis and therapy applications.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑