拟人呼吸胸腔体模Casper的设计和CT成像

Design and CT imaging of Casper, an anthropomorphic breathing thorax phantom

Josie Laidlaw, Nicolas Earl, Nihal Shavdia, Rayna Davis, Sarah Mayer, Dmitri Karaman, Devon Richtsmeier, Pierre-Antoine Rodesch, Magdalena Bazalova-Carter

摘要:

这项工作的目标是建立一个拟人化的胸部体模能够呼吸运动的材料模仿人体组织的X射线成像应用。名为Casper的胸部体模由树脂(主体)、泡沫(肺)、发光聚乳酸(骨)和天然聚乳酸(放置在肺中的肿瘤)组成。通过在台式系统上进行光子计数计算机断层成像(CT)成像,评价了所有材料在生产前的X射线衰减特性。呼吸运动通过止转轭机构实现,隔膜运动频率为10 - 20 rpm,位移为1 - 2 cm。Casper是通过3D打印模具和肋骨制造的,并在复杂的过程中组装。然后使用临床CT扫描仪扫描最终体模,以评价材料CT数和肿瘤运动程度。Casper CT值接近于软组织(46 HU)、肋骨(125 HU)、肺(~ 840 HU)和肿瘤(~ 45 HU)的人体CT值。对于2cm的隔膜位移,最大肿瘤位移为0.7cm。在静态CT图像中准确评估了5个肿瘤体积,平均绝对误差为4.3%。在动态CT图像中,由于运动模糊,与真实体积的平均绝对差异为10.3%,因此较小肿瘤的肿瘤大小被低估或较大肿瘤的肿瘤大小被高估。

英文摘要:

The goal of this work was to build an anthropomorphic thorax phantom capable of breathing motion with materials mimicking human tissues in x-ray imaging applications. The thorax phantom, named Casper, was composed of resin (body), foam (lungs), glow polyactic acid (bones) and natural polyactic acid (tumours placed in the lungs). X-ray attenuation properties of all materials prior to manufacturing were evaluated by means of photon-counting computed tomography (CT) imaging on a table-top system. Breathing motion was achieved by a scotch-yoke mechanism with diaphragm motion frequencies of 10 - 20 rpm and displacements of 1 to 2 cm. Casper was manufactured by means of 3D printing of moulds and ribs and assembled in a complex process. The final phantom was then scanned using a clinical CT scanner to evaluate material CT numbers and the extent of tumour motion. Casper CT numbers were close to human CT numbers for soft tissue (46 HU), ribs (125 HU), lungs (-840 HU) and tumours (-45 HU). For a 2 cm diaphragm displacement the largest tumour displacement was 0.7 cm. The five tumour volumes were accurately assessed in the static CT images with a mean absolute error of 4.3%. Tumour sizes were either underestimated for smaller tumours or overestimated for larger tumours in dynamic CT images due to motion blurring with a mean absolute difference from true volumes of 10.3%.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑