能量阈值偏差计算器:一种基于物理模型的自适应校正方案用于光子计数CT
Energy-Threshold Bias Calculator: A Physics-Model Based Adaptive Correction Scheme for Photon-Counting CT
AI总结:
本文提出了一种基于物理模型的自适应校正方案,用于减少光子计数CT中的光谱不一致性,通过能量阈值偏差计算器实现高效且稳健的校正。
摘要:
基于光子计数探测器的计算机断层扫描(PCCT)近年来取得了显著进展。然而,光谱不一致性,即在检测每个能量 bin 的计数中出现的像素间变化,很容易导致环形或带状伪影以及 CT 重建图像的不准确性。本文提出了一种新的基于物理模型的方法,通过两个术语来校正光谱不一致性:(1)一个固定的光谱倾斜项(能量阈值无关的过滤函数),在给定的能量阈值下确定;(2)一个可变的能量阈值偏差项,可以通过我们的光谱模型在阈值变化时直接计算。在校准阶段计算出这两个术语后,它们将被纳入我们的光谱模型中,以自适应地生成光谱校正向量以及在需要时生成材料分解向量,逐像素处理 PCCT 投影数据。使用一组最小的参数,具有明确物理意义,这种能量阈值偏差计算器(ETB-Cal)具有计算效率高、实现稳健性和方便性的优点,无需在校准中使用X射线荧光材料。为了验证我们的方法,进行了数值模拟和使用多种假体的物理实验,均在台式PCCT系统上进行。初步结果表明,非均匀性显著减少,对于Gammex多能假体,从29.3到5.8HU(与无校正相比,相对地,8.3HU由一个涉及多项式的基于模型的方法实现,但不需要明确建模和计算能量阈值偏差,但需要更多校准数据),对于京都头部假体,从27.9到3.2HU。
英文摘要:
Photon-counting detector based computed tomography (PCCT) has greatly advanced in recent years. However, spectral inconsistency, referring to inter-pixel variations in detected counts per energy bin, can easily leads to ring or band artifacts and inaccuracies in CT reconstructed images. This work proposes a novel physics-model based method to correct for spectral inconsistency by modeling it through two terms: (1) a fixed spectral skew term (energy threshold-independent filtration function) determined at a given energy threshold, and (2) a variable energy-threshold bias term that can be directly calculated by using our spectral model as the threshold changes. After the two terms being computed out in the calibration stage, they will be incorporated into our spectral model to adaptively generate the spectral correction vectors as well as the material decomposition vectors if needed, pixel-by-pixel for PCCT projection data. Using a minimum set of parameters with explicit physics meaning, such an energy-threshold bias calculator (ETB-Cal) has advantages of computational efficiency, robustness in implementation, and convenience with no need of X-ray fluorescence materials in calibration. To validate our method, both numerical simulations and physical experiments using multiple phantoms were carried out on a tabletop PCCT system, with preliminary results showing a significant reduction in non-uniformity, from 29.3 to 5.8 HU for Gammex multi-energy phantom versus no correction (comparatively, 8.3 HU was achieved by a polynomial-involving model-based approach with no explicit modeling and calculating of energy threshold bias but more calibration data required), and from 27.9 to 3.2 HU for the Kyoto head phantom.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。