用于放射治疗中端到端测试的拟人多模态(CT/MRI)体模原型

An anthropomorphic multimodality (CT/MRI) phantom prototype for end-to-end tests in radiation therapy

Raya R. Gallas, Nora Hünemohr, Armin Runz, Nina I. Niebuhr, Oliver Jäkel, Steffen Greilich

摘要:

随着体外射束治疗的复杂性日益增加,所谓的“端到端”测试旨在涵盖从治疗计划到随访的所有步骤,以满足对质量保证的高要求。由于磁共振成像(MRI)在治疗过程中的重要性日益增加,并且已建立的体模(如Alderson头)不能用于这些测试,因此必须开发新的多模态体模。在这里,我们提出了这样一个可定制的多模态头部模型的可行性研究。我们使用一组患者CT图像作为拟人头部形状的基础。容器-由环氧树脂组成-使用快速原型(3D打印)生产。体模受体包括鼻腔、两个软组织体积和颅骨。此外,将球形肿瘤体积定位在中心。用磷酸二钾基颅骨替代物、琼脂糖凝胶和蒸馏水填充体积。肿瘤体积填充有含氧量正常的剂量测定凝胶。放射治疗的整个工作流程(包括临床质子照射)可成功应用于体模。CT测量显示Hounsfield单位与所有替代品的参考值一致。MRI对比度与体模材料的预期一致,但骨替代物提供了不期望的高信号强度。通过聚合凝胶剂量计的T2加权读数进行的剂量验证可以成功显示颗粒范围方面的正确计划交付。

英文摘要:

With the increasing complexity of external beam therapy, so-called "end-to-end" tests are intended to cover all steps from therapy planning to follow-up to fulfill the high demands on quality assurance. As magnetic resonance imaging (MRI) gains growing importance in the treatment process and established phantoms (such as the Alderson head) cannot be used for those tests, novel multimodality phantoms have to be developed. Here, we present a feasibility study for such a customizable multimodality head phantom. We used a set of patient CT images as the basis for the anthropomorphic head shape. The recipient - consisting of an epoxy resin - was produced using rapid prototyping (3D printing). The phantom recipient includes a nasal air cavity, two soft tissues volumes and cranial bone. Additionally a spherical tumor volume was positioned in the center. The volumes were filled with dipotassium phosphate-based cranial bone surrogate, agarose gel, and distilled water. The tumor volume was filled with normoxic dosimetric gel. The entire workflow of radiation therapy including a clinical proton irradiation could be successfully applied to the phantom. CT measurements revealed Hounsfield units agreeing with reference values for all surrogates. The MRI contrasts corresponded to the expectations on the phantom materials except for the bone surrogate providing an undesirably high signal intensity. The dose verification by T2-weighted read out of the polymerization gel dosimeter could successfully show a correct plan delivery in terms of particle range.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑