使用双层平板探测器的CBCT散射校正

CBCT scatter correction with dual-layer flat-panel detector

Xin Zhang, Jixiong Xie, Ting Su, Jiongtao Zhu, Han Cui, Yuhang Tan, Dongmei Xia, Hairong Zheng, Dong Liang, Yongshuai Ge

AI总结:

针对DL-FPD双能CBCT受散射影响的问题,提出基于线性能量信号模型的e-Grid快速散射估计与校正方法,实验证明其显著降低伪影和非均匀性并提升多物质分解精度。

摘要:

背景:近年来,基于双层平板探测器(DL-FPD)的双能锥形束CT(DE-CBCT)成像日益普及。然而,DE-CBCT的图像质量仍受到康普顿散射X射线光子的限制。目的:本研究旨在为基于DL-FPD的CBCT成像开发一种能量调制的散射校正方法。方法:在DL-FPD中,穿过物体的一部分X射线光子(主要是低能一次光子和散射光子)被探测器顶层捕获,而其余X射线光子(主要是高能一次光子和散射光子)由探测器底层收集。基于两组不同的低能和高能测量结果,对DL-FPD上的双能一次信号和散射信号近似建立了线性信号模型。利用该信号模型可快速估计X射线散射分布。研究开展了水模体的Monte Carlo(MC)模拟,以验证这一新开发的散射估计方法。此外,还进行了水模体、头部模体和腹部模体的物理实验,以验证所提出散射校正方法的实际性能。结果:MC结果表明,e-Grid方法能够生成接近真值的散射分布。此外,物理实验证明,e-Grid方法可大幅减少DL-FPD获取的低能和高能CBCT图像中的阴影伪影。平均而言,低能CBCT图像的图像非均匀性(NU)降低超过77%,高能CBCT图像降低超过66%。因此,分解得到的多物质基的准确性得到显著提升。

英文摘要:

Background: Recently, the popularity of dual-layer flat-panel detector (DL-FPD) based dual-energy cone-beam CT (DE-CBCT) imaging has been increasing. However, the image quality of DE-CBCT remains constrained by the Compton scattered X-ray photons. Purpose: The objective of this study is to develop an energy-modulated scatter correction method for DL-FPD based CBCT imaging. Methods: In DL-FPD, a certain portion of the X-ray photons (mainly low-energy primary and scattered photons) passing through the object are captured by the top detector layer, while the remaining X-ray photons (mainly high-energy primary and scattered photons) are collected by the bottom detector layer. Based on the two set of distinct low-energy and high-energy measurements, a linear signal model was approximated for the dual-energy primary and scattered signals on DL-FPD. The distributions of X-ray scatters were quickly estimated using this signal model. Monte Carlo (MC) simulation of a water phantom was conducted to verify this newly developed scatter estimation method. Moreover, physical experiments of water phantom, head phantom, and abdominal phantom were carried out to validate the real performance of this proposed scatter correction method. Results: The MC results showed that the e-Grid method was able to generate scatter distributions close to the ground truth. Moreover, the physical experiments demonstrated that the e-Grid method can greatly reduce the shading artifacts in both low-energy and high-energy CBCT images acquired from DL-FPD. On average, the image non-uniformity (NU) was reduced by over 77% in the low-energy CBCT image and by over 66% in the high-energy CBCT image. A a consequence, the accuracy of the decomposed multi-material bases was substantially improved.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑