arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1503.03527physics.med-ph

Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study

Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu Ju, Yoonsun Chung, Sang Hoon Jung, Youngyih Han

英文摘要:

Two full rotating gantry with different nozzles (Multipurpose nozzle with MLC, Scanning Dedicated nozzle) with conventional cyclotron system is installed and under commissioning for various proton treatment options at Samsung Medical Center in Korea. The purpose of this study is to investigate neutron dose equivalent per therapeutic dose, H/D, to x-ray imaging equipment under various treatment conditions with monte carlo simulation. At first, we investigated H/D with the various modifications of the beam line devices (Scattering, Scanning, Multi-leaf collimator, Aperture, Compensator) at isocenter, 20, 40, 60 cm distance from isocenter and compared with other research groups. Next, we investigated the neutron dose at x-ray equipments used for real time imaging with various treatment conditions. Our investigation showed the 0.07 ~ 0.19 mSv/Gy at x-ray imaging equipments according to various treatment options and intestingly 50% neutron dose reduction effect of flat panel detector was observed due to multi- leaf collimator during proton scanning treatment with multipurpose nozzle. In future studies, we plan to investigate experimental measurement of neutron dose and validation of simulation data for x-ray imaging equipment with additional neutron dose reduction method.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑