基于Monte Carlo方法的锥束CT系统模拟
Simulation of Cone Beam CT System Based on Monte Carlo Method
摘要:
自适应放射治疗(ART)是在图像引导放射治疗(IGRT)的基础上发展起来的,是光子放射治疗的发展趋势。为了更好地利用锥形束CT(CBCT)图像进行ART,基于Monte Carlo程序建立了CBCT系统模型,并进行了测量验证。KV X射线管采用BEAMnrc程序。选择IOURCE-13和ISOURCE-24来模拟束流粒子的路径。用DOSXYZnrc程序计算了1cm水下的剂量分布,并与实测的深度剂量分布和侧向剂量分布进行了比较。在10 cm深度范围内,计算的PDD优于1%。计算的侧向剂量分布中超过85%的点在2%以内。正确的CBCT系统模型有助于提高ART剂量验证的CBCT图像质量,并评估CBCT图像伴随的剂量风险。
英文摘要:
Adaptive Radiation Therapy (ART) was developed based on Image-guided Radiation Therapy (IGRT) and it is the trend of photon radiation therapy. To get a better use of Cone Beam CT (CBCT) images for ART, the CBCT system model was established based on Monte Carlo program and validated against the measurement. The BEAMnrc program was adopted to the KV x-ray tube. Both IOURCE-13 and ISOURCE-24 were chosen to simulate the path of beam particles. The measured Percentage Depth Dose (PDD) and lateral dose profiles under 1cm water were compared with the dose calculated by DOSXYZnrc program. The calculated PDD was better than 1% within the depth of 10cm. More than 85% points of calculated lateral dose profiles was within 2%. The correct CBCT system model helps to improve CBCT image quality for dose verification in ART and assess the CBCT image concomitant dose risk.