剂量率为1000的临床笔形束扫描质子治疗装置的调试(FLASH)
Commissioning of a clinical pencil beam scanning proton therapy unit for ultrahigh dose rates (FLASH)
摘要:
目的:这项工作的目的是提供一个灵活的平台,通过调整以前的临床笔形束扫描机架的照射剂量率与质子的FLASH研究。方法:PSI机架1治疗患者直至2018年12月。我们优化了光束线参数,以将从PSI COMET加速器提取的250 MeV光束传输到治疗室,最大限度地提高光束强度到样品的传输。我们的特点是剂量监测器的机架上,以确保良好的控制剂量,提供点扫描模式。我们的特点是不同的剂量率和字段大小的传输辐照的光束。我们探索了扫描的可能性,以使大目标的共形辐射或透射辐射(横向扫描)。结果:我们实现了从回旋加速器到治疗室的86%的传输。我们达到了9000 Gy/s的峰值剂量率在3毫米水当量深度,沿一个单一的笔形束的中心轴。单点闪光照射可实现高达5 × 5 mm $^{2}$的射野尺寸。快速横向扫描可覆盖16 x 1.2 cm $^{2}$的视野。通过使用安装在桅杆上的量程转换器,我们能够在19.6至37.9厘米的水中跨越深度。输送的各种剂量水平的精度小于1%。结论:我们已经实现了一个质子FLASH照射设置能够连续调查宽剂量率谱,从1到9000 Gy/s的单点照射,以及在笔形束扫描模式。因此,我们已经开发了一个多功能的测试平台,闪存的研究。
英文摘要:
Purpose: The purpose of this work was to provide a flexible platform for FLASH research with protons by adapting a former clinical pencil beam scanning gantry to irradiations with ultrahigh dose rates. Methods: PSI Gantry 1 treated patients until December 2018. We optimized the beamline parameters to transport the 250 MeV beam extracted from the PSI COMET accelerator to the treatment room, maximizing the transmission of beam intensity to the sample. We characterized a dose monitor on the gantry to ensure good control of the dose, delivered in spot-scanning mode. We characterized the beam for different dose rates and field sizes for transmission irradiations. We explored scanning possibilities in order to enable conformal irradiations or transmission irradiations of large targets (with transverse scanning). Results: We achieved a transmission of 86 % from the cyclotron to the treatment room. We reached a peak dose rate of 9000 Gy/s at 3 mm water equivalent depth, along the central axis of a single pencil beam. Field sizes of up to 5x5 mm$^{2}$ were achieved for single spot FLASH irradiations. Fast transverse scanning allowed to cover a field of 16x1.2 cm$^{2}$. With the use of a nozzle-mounted range shifter we are able to span depths in water ranging from 19.6 to 37.9 cm. Various dose levels were delivered with a precision within less than 1 %. Conclusions: We have realized a proton FLASH irradiation setup able to investigate continuously a wide dose rate spectrum, from 1 to 9000 Gy/s in a single spot irradiation as well as in the pencil beam scanning mode. As such, we have developed a versatile test bench for FLASH research.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。