Multi-Energy Blended CBCT Spectral Imaging Using a Spectral Modulator with Flying Focal Spot (SMFFS)

Yifan Deng, Hao Zhou, Zhilei Wang, Adam S. Wang, Hewei Gao

AI总结:

本文首次提出带飞焦点的光谱调制器(SMFFS)技术,并设计散射解耦材料分解(SDMD)方法,在统一框架中同时实现CBCT光谱成像与散射校正,显著提升材料分解精度和CT定量成像性能。

摘要:

传统锥形束CT(CBCT)容易受到散射和射束硬化伪影的影响,而散射与光谱效应的纠缠进一步增加了复杂性。在本工作中,我们首次尝试开发一种静态的带飞焦点的光谱调制器(SMFFS)技术,作为一种有前景的低成本方法,在统一框架中精确求解X射线散射问题并在物理上实现光谱成像。为应对散射与光谱相互交织的挑战,我们基于散射相似性假设,提出了一种面向SMFFS的新型散射解耦材料分解(SDMD)方法。对不同焦点偏转下的纯水圆柱体模的蒙特卡洛模拟表明,在约2 mm范围内的焦点偏转总体上具有非常相似的散射分布。使用临床腹部CT数据集的数值仿真表明,结合SDMD方法的SMFFS能够实现更好的材料分解和CT值精度,且伪影更少。在桌面CBCT系统上使用Gammex多能CT体模和拟人胸部体模进行的物理实验验证了SMFFS实现CBCT光谱成像的可行性。对于胸部体模,在70 keV虚拟单能图像(VMI)的选定感兴趣区域(ROIs)中,SMFFS CB扫描的均方根误差(RMSE)为11.8 HU,而顺序80/140 kVp(DKV)CB扫描在有散射校正和无散射校正时分别为14.5 HU和437.6 HU。此外,选定区域间的非均匀性在SMFFS CB扫描中为14.1 HU,而采用传统散射校正方法和未采用传统散射校正方法的DKV CB扫描分别为59.4 HU和184.0 HU。我们的初步结果表明,SMFFS能够在CBCT中实现光谱成像并同时进行散射校正,有效提升其定量成像性能。

英文摘要:

Conventional cone-beam CT (CBCT) can be easily compromised by scatter and beam hardening artifacts, and the entanglement of scatter and spectral effects introduces additional complexity. In this work, we present the first attempt to develop a stationary spectral modulator with flying focal spot (SMFFS) technology as a promising, low-cost approach to accurately solving the X-ray scattering problem and physically enabling spectral imaging in a unified framework. To deal with the intertwined scatter-spectral challenge, we propose a novel scatter-decoupled material decomposition (SDMD) method for SMFFS based on a hypothesis of scatter similarity. Monte Carlo simulations of a pure-water cylinder phantom with different focal spot deflections show that focal spot deflections within a range of ~2 mm share quite similar scatter distributions overall. Numerical simulations using a clinical abdominal CT dataset demonstrate that SMFFS with SDMD method can achieve better material decomposition and CT number accuracy with less artifacts. Physics experiments on a tabletop CBCT system using a Gammex multi-energy CT phantom an anthropomorphic chest phantom, are carried out to demonstrate the feasibility of CBCT spectral imaging with SMFFS. For the chest phantom, the root mean square error (RMSE) in selected regions of interest (ROIs) of virtual monochromatic image (VMI) at 70 keV is 11.8 HU for SMFFS CB scan, and 14.5 and 437.6 HU for sequential 80/140 kVp (DKV) CB scan with and without scatter correction, respectively. Also, the non-uniformity among selected regions is 14.1 HU for SMFFS CB scan, and 59.4 and 184.0 HU for the DKV CB scan with and without a traditional scatter correction method, respectively. Our preliminary results show that SMFFS can enable spectral imaging with simultaneous scatter correction for CBCT and effectively improve its quantitative imaging performance.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑