Fastcat kV和MV锥束CT(CBCT)模拟器的实验验证

Experimental validation of Fastcat kV and MV cone beam CT (CBCT) simulator

Jericho O'Connell, Clayton Lindsay, Magdalena Bazalova-Carter

摘要:

目的:针对Varian Truebeam直线加速器采集的kV和MV CBCT图像,对Fastcat锥形束CT(CBCT)模拟器进行实验验证。研究方法:Catphan 504体模的kV和MV CBCT图像分别使用机载成像仪(OBI)的100 kVp射束和电子射野成像设备(EPID)的6 MV治疗射束采集。使用X射线源、蝴蝶结滤波器、高分辨率体素化虚拟Catphan体模、防散射滤线栅和CsI闪烁探测器的详细模型进行kV Fastcat模拟。同样,使用射束能谱、平坦滤波器、高分辨率体素化虚拟Catphan体模和GOS闪烁探测器的详细模型模拟MV Fastcat CBCT。比较了体模的实验和模拟CBCT图像的HU值、对比噪声比(CNR)和剂量线性。探测器调制传递函数(MTF)的两个探测器也进行了实验验证。将Fastcat的剂量计算与使用Topas进行的MC剂量计算进行比较。结果:对于kV和MV模拟,对比度分别在14和9 HU范围内,探测器MTF分别在4.2%和2.5%范围内。同样,CNR的均方根误差(RMSE)为2.6%和1.4%。剂量在MC值的2.4%和1.6%范围内一致。kV和MV CBCT图像在Fastcat中分别用887和493个投影进行模拟需要71和72秒。结论:我们提出了一个多能量的实验验证的快速和准确的CBCT模拟器对商业直线加速器。该模拟器是开源的,在这项工作中发现的所有模型都可以从https://github.com/jerichooconnell/fastcat.git下载

英文摘要:

Purpose: To experimentally validate the Fastcat cone beam CT (CBCT) simulator against kV and MV CBCT images acquired with a Varian Truebeam linac. Methods: kV and MV CBCT images of a Catphan 504 phantom were acquired using a 100 kVp beam with the on-board imager (OBI) and a 6 MV treatment beam with the electronic portal imaging device (EPID), respectively. The kV Fastcat simulation was performed using detailed models of the x-ray source, bowtie filter, a high resolution voxelized virtual Catphan phantom, anti-scatter grid, and the CsI scintillating detector. Likewise, an MV Fastcat CBCT was simulated with detailed models for the beam energy spectrum, flattening filter, a high resolution voxelized virtual Catphan phantom, and the GOS scintillating detector. Experimental and simulated CBCT images of the phantom were compared with respect to HU values, contrast to noise ratio (CNR),and dose linearity. Detector modulation transfer function (MTF) for the two detectors were also experimentally validated. Fastcat's dose calculations were compared to MC dose calculations performed with Topas. Results: For the kV and MV simulations, respectively: Contrast agreed within 14 and 9 HUs and detector MTF agreed within 4.2% and 2.5%. Likewise, CNR had a root mean squared error (RMSE) of 2.6% and 1.4%. Dose agreed within 2.4% and 1.6% of MC values. The kV and MV CBCT images took 71 and 72 seconds to simulate in Fastcat with 887 and 493 projections, respectively. Conclusions: We present a multi energy experimental validation of a fast and accurate CBCT simulator against a commercial linac. The simulator is open source and all models found in this work can be downloaded from https://github.com/jerichooconnell/fastcat.git

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑