千伏CBCT成像和兆伏射束输送的同时进行:使用2D防散射滤线栅和基于滤线栅的散射采样抑制交叉散射

Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: Suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling

Farhang Bayat, Mohamed Elsayed Eldib, Brian Kavanagh, Moyed Miften, Cem Altunbas

摘要:

同时使用千伏(kV)和兆伏(MV)射束的概念在锥形射束计算机断层成像(CBCT)引导的放射治疗中具有潜在应用,例如单次屏气扫描、金属伪影减少和MV治疗递送期间的同时成像。然而,MV射束输送期间产生的MV交叉散射降低了CBCT图像质量。为了解决这一问题,研究了2D抗散射网格和交叉散射校正方法。利用3D打印的钨2D抗散射栅格原型来减少在同时MV射束输送期间kV投影中的MV交叉散射注量。通过使用2D网格本身作为交叉散射强度采样装置来校正剩余的交叉散射,称为基于网格的散射采样。为了测试这种方法,在输送6和10 MV射束的同时进行kV CBCT采集,模拟高剂量率治疗输送场景。在体模的投影和CBCT图像中评估了所提出的方法的MV交叉散射抑制性能。2D网格将kV投影中MV交叉散射的强度平均降低了3倍。通过基于网格的散射采样估计的剩余MV交叉散射在测量的参考强度值的7%以内。CBCT图像质量在同步kV-MV射束输送过程中得到显著改善。在没有我们的方法的情况下,Hounsfield单位(HU)的中位不准确度高达191 HU,并且使用我们的2D网格和散射校正方法将其降低到3 HU。我们的方法提供了2-6倍的对比度比改善。结果表明,我们的方法可以成功地减少高能量交叉散射的影响,同时kV CBCT成像和兆伏治疗交付。

英文摘要:

The concept of using kilovoltage (kV) and megavoltage (MV) beams concurrently has potential applications in cone beam computed tomography (CBCT) guided radiation therapy, such as single breath hold scans, metal artifact reduction, and simultaneous imaging during MV treatment delivery. However, MV cross-scatter generated during MV beam delivery degrades CBCT image quality. To address this, a 2D antiscatter grid and cross scatter correction method were investigated. a 3D printed, tungsten 2D antiscatter grid prototype was utilized to reduce MV cross-scatter fluence in kV projections during concurrent MV beam delivery. Remaining cross-scatter was corrected by using the 2D grid itself as a cross-scatter intensity sampling device, referred as Grid-based Scatter Sampling. To test this approach, kV CBCT acquisitions were performed while delivering 6 and 10 MV beams, mimicking high dose rate treatment delivery scenarios. MV cross-scatter suppression performance of the proposed approach was evaluated in projections and CBCT images of phantoms. 2D grid reduced the intensity of MV cross-scatter in kV projections by a factor of 3 on the average. Remaining MV cross-scatter estimated by Grid-based Scatter Sampling was within 7% of measured reference intensity values. CBCT image quality was improved substantially during concurrent kV-MV beam delivery. Median Hounsfield Unit (HU) inaccuracy was up to 191 HU without our methods, and it was reduced to 3 HU with our 2D grid and scatter correction approach. Our methods provided a factor of 2-6 improvement in contrast-to-ratio. Results indicate that our approach can successfully minimize the effects of high energy cross-scatter in concurrent kV CBCT imaging and megavoltage treatment delivery.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑