Single-scan scatter correction in CBCT by using projection correlation based view interpolation (PC-VI) and a stationary ring-shaped beam stop array (BSA)

Hao Yan, Xuanqin Mou, Yanbo Zhang, Maria Zankl

摘要:

在X射线锥形束(CB)CT的散射校正中,使用移动束流阻挡阵列(BSA)的单次扫描方案提供了可靠的低剂量散射测量,并且通过使用基于投影相关的视图插值(PC-VI),可以高精度地恢复(在散射测量期间)被移动BSA遮蔽的主要通量。然而,在实际应用中,移动的BSA可能增加机械负担。为了更好的实用性,本文提出了一种基于PC-VI的单次扫描方案,其中环形固定BSA作为CB扫描过程中的虚拟移动BSA,因此该固定BSA遮蔽的初级通量也可以通过PC-VI很好地恢复。对整个系统的设计原则进行了推导和评价。该方案大大增强了单扫描散射校正方案的实用性。

英文摘要:

In the scatter correction for x-ray Cone Beam (CB) CT, the single-scan scheme with moving Beam Stop Array (BSA) offers reliable scatter measurement with low dose, and by using Projection Correlation based View Interpolation (PC-VI), the primary fluence shaded by the moving BSA (during scatter measurement) could be recovered with high accuracy. However, the moving BSA may increase the mechanical burden in real applications. For better practicability, in this paper we proposed a PC-VI based single-scan scheme with a ring-shaped stationary BSA, which serves as a virtual moving BSA during CB scan, so the shaded primary fluence by this stationary BSA can be also well recovered by PC-VI. The principle in designing the whole system is deduced and evaluated. The proposed scheme greatly enhances the practicability of the single-scan scatter correction scheme.