组织T2弛豫时间与体内组织扩散测量之间的复杂关系,取决于T2值的范围

Complicated relationships between tissue T2 relaxation time and in vivo tissue diffusion measures, depending on the ranges of T2 value

Yi Xiang J. Wang

AI总结:

本文通过文献证据论证ADC值受T2弛豫影响,指出在特定T2范围内,ADC测量主要由T2效应主导而非扩散信息。

摘要:

表观扩散系数(ADC)是衡量组织内水分子扩散程度的指标。我们认为ADC值同时包含扩散和T2弛豫的信息。在本信中,我们列举文献证据支持这一论点。首先,我们以子宫肌层肿瘤为例。肌层在3T下的T2弛豫时间为79毫秒。文献显示,当肌层肿瘤的T2加权信号强度相对于肌层呈高信号时,这些肿瘤可能表现出扩散受限。这类似于脾脏相对于肝脏呈T2加权高信号,而脾脏相对于肝脏表现出扩散受限。另一方面,当肌层肿瘤的T2加权信号强度相对于肌层呈低信号时,这些肿瘤可能不表现出扩散受限。然而,当肌层肿瘤呈极高信号时,如平滑肌瘤囊性变和黏液样变的病例,T2加权信号强度与扩散之间的关系类似于正常胆囊,即T2加权高信号而无扩散受限。对于大多数肿瘤,较长的T2可能提示肿瘤含水量更高。在b=0图像上,肿瘤(相对于其他组织)的信号可能由T2效应主导。在高b值图像(如b=1000)上,肿瘤区域可能由噪声主导。最终,由b=0信号与b=1000信号之间的斜率确定的ADC值也受T2效应主导,较长的T2与较高的ADC测量值相关。实际上,由于b=0图像没有扩散梯度,而b=1000(或b=800)图像仅有噪声,ADC图缺乏扩散信息。

英文摘要:

Apparent diffusion coefficient (ADC) is a measure of the magnitude of diffusion of water molecules within tissues. We argue that ADC value contains information of both diffusion and T2 relaxation. In this letter, we list literature evidence to support our argument. Firstly, we list uterine myometrium tumors as examples. Myometrium has a T2 relaxation time of 79 ms at 3T. Literature shows, when the myometrium tumors are T2 weighted signal intensity hypertensive (relative to myometrium), these tumors likely show diffusion restriction. This is similar to that, while the spleen is T2 weighted signal intensity hypertensive relative to the liver, the spleen demonstrates diffusion restriction relative to the liver. On the other hand, when the myometrium tumors are T2 weighted signal intensity hypotensive (relative to myometrium), these tumors likely do not show diffusion restriction. However, when the myometrium tumors are very highly hypertensive such as the cases of leiomyoma cystic degeneration and myxoid degeneration, the relationship between T2 weighted signal intensity and diffusion is similar to that of a normal gallbladder, i.e., T2 weighted signal highly hypertensive without diffusion restriction. For most of the tumors, much longer T2 may suggest a tumor is waterier. On b=0 images, the signal of tumor (relative to other tissues) can be dominated by T2 effect. On a high b-value images such as b=1000 images, it is possible that tumor regions are dominated by noises. In the end, the ADCs, which are determined by the slope between the signal at b=0 and the signal at b=1000, are also dominated by the T2 effect, with longer T2 being associated with higher ADC measure. In fact, since there is no diffusion gradient with b=0 images, and there are only noises on b=1000 (or b=800), ADC map is devoid of diffusion information.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑