首次使用模块化J-PET扫描仪用$^{44}$Sc放射性核素进行体外正电子对成像

First ex-vivo positronium imaging of tissues with modular J-PET scanner using $^{44}$Sc radionuclide

Jagiellonian University(雅盖隆大学);AGH University of Science and Technology(AGH科技大学);Institute of Nuclear Chemistry and Technology(核化学与技术研究所);University of Warsaw(华沙大学);Cracow University of Technology(克拉科夫理工大学);INFN, Laboratori Nazionali di Frascati(意大利国家核物理研究所,弗拉斯卡蒂国家实验室)
Karol Kubat, Manish Das, Sushil Sharma, Ermias Y. Beyene, Aleksander Bilewicz, Jarosław Choiński, Neha Chug, Catalina Curceanu, Eryk Czerwiński, Jakub Hajduga, … 展开作者

Karol Kubat, Manish Das, Sushil Sharma, Ermias Y. Beyene, Aleksander Bilewicz, Jarosław Choiński, Neha Chug, Catalina Curceanu, Eryk Czerwiński, Jakub Hajduga, Sharareh Jalali, Krzysztof Kacprzak, Tevfik Kaplanoglu, Łukasz Kapłon, Kamila Kasperska, Aleksander Khreptak, Grzegorz Korcyl, Tomasz Kozik, Deepak Kumar, Sumit Kumar Kundu, Anoop Kunimmal-Venadan, Bartosz Leszczyński, Edward Lisowski, Filip Lisowski, Justyna Mędrala-Sowa, Simbarashe Moyo, Wiktor Mryka, Szymon Niedźwiecki, Anand Pandey, Piyush Pandey, Szymon Parzych, Alessio Porcelli, Bartłomiej Rachwał, Martin Rädler, Magdalena Skurzok, Anna Stolarz, Tomasz Szumlak, Pooja Tanty, Keyvan Tayefi Ardebili, Satyam Tiwari, Kavya Valsan Eliyan, Rafał Walczak, Paweł Moskal, Ewa Ł. Stępień

AI总结:

本研究首次使用模块化J-PET扫描仪和$^{44}$Sc放射性核素进行体外正电子对成像,验证了其在正电子对寿命成像中的应用潜力。

摘要:

本研究首次使用模块化J-PET扫描仪用$^{44}$Sc放射性核素进行人体组织的体外正电子对成像。$^{44}$Sc同位素通过$^{44}$Ca(p, n)$^{44}$Sc核反应产生,用于对由人体脂肪组织、心肌瘤组织、血栓血凝块以及多孔聚合物XAD4和由石英熔融制成的认证参考材料(CRM)组成的假体进行正电子对成像。实验证明了$^{44}$Sc作为正电子对成像的正电子源的适用性。J-PET对$^{44}$Sc进行正电子对成像的性能通过正确重建CRM材料和XAD-4聚合物的平均正电子对寿命得到验证。确定的脂肪组织、心肌瘤组织和血栓的平均正电子对(oPs)寿命与先前实验结果一致。本研究突显了$^{44}$Sc放射性核素在正电子对寿命成像(PLI)中的潜力。

英文摘要:

This study presents the first ex-vivo positronium imaging of human tissues using the modular J-PET scanner with the $^{44}$Sc radionuclide. The $^{44}$Sc isotope was produced via the $^{44}$Ca(p, n)$^{44}$Sc nuclear reaction and used to perform positronium imaging of phantom composed of human adipose tissue, cardiac myxoma tissue, thrombi blood clot, and also porous polymer XAD4, and a certified reference material (CRM) made from fused silica. The experiment demonstrates the suitability of $^{44}$Sc as a positron source for positronium imaging. The performance of J-PET for positronium imaging with $^{44}$Sc was validated by proper reconstruction of the mean orthopositronium lifetime for CRM material and XAD-4 polymer. The mean ortho-positronium (oPs) lifetimes determined for adipose tissue, cardiac myxoma tissues and thrombi were consistent with results of previous experiments. The study highlights the potential $^{44}$Sc radionuclide for positronium lifetime imaging (PLI).