使用124I在Biograph Vision Quadra上进行正电子素寿命成像

Positronium Lifetime Imaging with the Biograph Vision Quadra using 124I

Lorenzo Mercolli, William M. Steinberger, Narendra Rathod, Maurizio Conti, Paweł Moskal, Axel Rominger, Robert Seifert, Kuangyu Shi, Ewa Ł. Stępień, Hasan Sari

AI总结:

本研究利用$^{124}$I和商用LAFOV PET/CT,通过Amberlite XAD4体模验证体素级正-正电子素寿命成像的可行性及空间区分能力。

摘要:

目的:测量人体组织中的正-正电子素(oPs)寿命,有可能为传统正电子发射断层扫描(PET)增加有关组织微环境的临床相关信息。通过体模测量,我们研究了使用商用长轴向视野(LAFOV)PET扫描仪进行oPs寿命的体素级测量。方法:我们制备了四个由多孔聚合物吸附剂Amberlite XAD4与水混合而成的样品,并加入1.12 MBq至1.44 MBq的$^{124}$I。样品在两种不同设置下进行扫描:一次各样品之间相隔数厘米(扫描时间15分钟),另一次将所有样品捆扎在一起(扫描时间40分钟)。对于每次扫描,我们测定完整样品以及体素层面的oPs寿命。所考察的体素尺寸为$10.0^3$ mm$^3$、$7.1^3$ mm$^3$和$4.0^3$ mm$^3$。结果:Amberlite XAD4能够制备具有不同oPs寿命的样品。采用贝叶斯拟合程序,完整样品中的oPs寿命分别为$2.52 \pm 0.03$ ns、$2.37\pm0.03$ ns、$2.27\pm0.04$ ns和$1.82\pm0.02$ ns。体素级oPs寿命拟合表明,即使采用$4.0^3$ mm$^3$体素,各样品仍可清晰区分,且中心体素具有良好的计数统计。然而,在样品紧密相邻的情况下,对具有不同oPs寿命区域进行空间区分仍具挑战性。结论:我们的研究表明,使用$^{124}$I在商用LAFOV PET/CT上进行正电子素寿命成像,在临床条件下应当是可行的。

英文摘要:

Purpose: Measuring the ortho-positronium (oPs) lifetime in human tissue bears the potential of adding clinically relevant information about the tissue microenvironment to conventional positron emission tomography (PET). Through phantom measurements, we investigate the voxel-wise measurement of oPs lifetime using a commercial long-axial field-of-view (LAFOV) PET scanner. Methods: We prepared four samples with mixtures of Amberlite XAD4, a porous polymeric adsorbent, and water and added between 1.12 MBq and 1.44 MBq of $^{124}$I. The samples were scanned in two different setups: once with a couple of centimeters between each sample (15 minutes scan time) and once with all samples taped together (40 minutes scan time). For each scan, we determine the oPs lifetime for the full samples and at the voxel level. The voxel sizes under consideration are $10.0^3$ mm$^3$, $7.1^3$ mm$^3$ and $4.0^3$ mm$^3$. Results: Amberlite XAD4 allows the preparation of samples with distinct oPs lifetime. Using a Bayesian fitting procedure, the oPs lifetimes in the whole samples are $2.52 \pm 0.03$ ns, $2.37\pm 0.03$ ns, $2.27\pm0.04$ ns and $1.82\pm 0.02$ ns, respectively. The voxel-wise oPs lifetime fits showed that even with $4.0^3$ mm$^3$ voxels the samples are clearly distinguishable and a central voxels have good count statistics. However, the situation with the samples close together remains challenging with respect to the spatial distinction of regions with different oPs lifetimes. Conclusion: Our study shows that positronium lifetime imaging on a commercial LAFOV PET/CT should be feasible under clinical conditions using $^{124}$I.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑