arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2011.09698physics.bio-ph
收藏

向细胞和组织培养物传递感应电磁场的新型体外装置

A novel in vitro device to deliver induced electromagnetic fields to cell and tissue cultures

Rea Ravin, Teddy X. Cai, Randall H. Pursley, Marcial Garmendia-Cedillos, Tom Pohida, Raisa Z. Freidlin, Herui Wang, Zhengping Zhuang, Amber J. Giles, Nathan H. … 展开作者

Rea Ravin, Teddy X. Cai, Randall H. Pursley, Marcial Garmendia-Cedillos, Tom Pohida, Raisa Z. Freidlin, Herui Wang, Zhengping Zhuang, Amber J. Giles, Nathan H. Williamson, Mark R. Gilbert, Peter J. Basser

摘要:

我们已经开发了一种新型的体外仪器,可以提供中频(100 - 400 kHz),中等强度(高达和超过6.5 V/cm峰峰值)电场(EF)的细胞和组织培养使用电磁场(EMF)产生的螺线管线圈。这些EF的主要应用是作为一种新兴的癌症治疗方式。Novocure Ltd.的体外研究报告称,中频(100 - 300 kHz)、低振幅(1 - 3 V/cm)EF(他们称之为“肿瘤治疗场”(TTFields))对多形性胶质母细胞瘤(GBM)细胞具有抗有丝分裂作用。发现该效果随着EF振幅的增加而增加。尽管持续的理论、临床前和临床研究,但作用机制仍不完全清楚。之前的“TTFields”体外研究使用了连接的电容耦合电极,将交变电场输送至细胞和组织培养物。这种接触递送方法的缺点是表征不佳的EF曲线和传导加热,这限制了所施加的EF的持续时间和幅度。相比之下,我们的设备以非接触方式输送具有良好特征的径向轮廓的EF,消除了传导加热并实现了热调节EF输送。为了测试和证明我们的系统,我们产生了连续的200 kHz EMF,EF振幅分布范围为0 - 6.5 V/cm pk-pk,并将其应用于人甲状腺细胞培养72小时。我们观察到在低EF振幅(< 4 V/cm)下细胞密度中度降低(< 10%),在较高振幅(4 - 6.5 V/cm)下细胞密度降低更大,高达25%。我们的设备可以扩展到其他EF频率和幅度制度。该器械的未来研究应通过更好地隔离EF的作用,促进关于“TTFields”有效性和作用机制的持续辩论。

英文摘要:

We have developed a novel in vitro instrument that can deliver intermediate frequency (100 - 400 kHz), moderate intensity (up to and exceeding 6.5 V/cm pk-pk) electric fields (EFs) to cell and tissue cultures generated using induced electromagnetic fields (EMFs) in a solenoid coil. A major application of these EFs is as an emerging cancer treatment modality. In vitro studies by Novocure Ltd. reported that intermediate frequency (100 - 300 kHz), low amplitude (1 - 3 V/cm) EFs, which they called "Tumor Treating Fields" (TTFields), had an anti-mitotic effect on glioblastoma multiforme (GBM) cells. The effect was found to increase with increasing EF amplitude. Despite continued theoretical, preclinical, and clinical study, the mechanism of action remains incompletely understood. Previous in vitro studies of "TTFields" have used attached, capacitively coupled electrodes to deliver alternating EFs to cell and tissue cultures. This contacting delivery method suffers from a poorly characterized EF profile and conductive heating that limits the duration and amplitude of the applied EFs. In contrast, our device delivers EFs with a well-characterized radial profile in a non-contacting manner, eliminating conductive heating and enabling thermally regulated EF delivery. To test and demonstrate our system, we generated continuous 200 kHz EMF with an EF amplitude profile spanning 0 - 6.5 V/cm pk-pk and applied them to human thyroid cell cultures for 72 hours. We observed moderate reduction in cell density (< 10%) at low EF amplitudes (< 4 V/cm) and a greater reduction in cell density of up to 25% at higher amplitudes (4 - 6.5 V/cm). Our device can be extended to other EF frequency and amplitude regimes. Future studies with this device should contribute to the ongoing debate about the efficacy and mechanism(s) of action of "TTFields" by better isolating the effects of EFs.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑