放射性分析用HPGe探测器的GEANT4模型

GEANT4 models of HPGe detectors for radioassay

R. H. M. Tsang, A. Piepke, D. J. Auty, B. Cleveland, S. Delaquis, T. Didberidze, R. MacLellan, Y. Meng, O. Nusair, T. Tolba

摘要:

在GEANT 4 \cite{geant4}中建立了位于阿拉巴马大学的两个高纯锗探测器GeII和GeIII的辐射输运模型。这些探测器已被广泛用于各种低本底实验中所用材料的放射性分析测量。这两个模型已经验证了对实际数据的几种情况下,通常看到在放射性测定测量。GeII和GeIII模型的系统不确定度分别为$\sim$ 12%和$\sim$ 9%。

英文摘要:

Radiation transport models of two high purity germanium detectors, GeII and GeIII, located at the University of Alabama have been created in GEANT4 \cite{geant4}. These detectors have been used extensively for radioassay measurements of materials used in various low background experiments. The two models have been validated against actual data under several scenarios typically seen in radioassay measurements. The systematic uncertainties of the models for GeII and GeIII are estimated to be $\sim$12\% and $\sim$9\% respectively.