HPGe探测器能量分辨率与数字化仪采样性能关系的分析与验证
Analysis and Verification of Relation between Digitizer's Sampling Properties and Energy Resolution of HPGe Detectors
摘要:
CDEX(中国暗物质实验)旨在探测76 Ge的弱相互作用大质量粒子(WIMP)和无中子双β衰变(0 vbb)。它现在在CJPL(中国锦屏地下实验室)使用~10 kg HPGe(高纯锗)探测器。通过6 μ s整形时间波形的高度谱计算了探测器的能量分辨率。有必要了解数字化仪的采样特性如何影响能量分辨率。本文将提出初步的能量分辨率结果的波形在不同的采样属性。初步结果表明,有效位数为8.25位或更高的探测器可以满足CDEX HPGe探测器@122keV的能量分辨率要求。基于ADC量化误差理论,本文还将对CDEX HPGe探测器的能量分辨率进行定量分析。这将为HPGe探测器全链低温读出电子学中ADC的设计提供指导。
英文摘要:
The CDEX (China Dark matter Experiment) aims at detection of WIMPs (Weakly Interacting Massive Particles) and 0vbb (Neutrinoless double beta decay) of 76Ge. It now uses ~10 kg HPGe (High Purity Germanium) detectors in CJPL (China Jinping Underground Laboratory). The energy resolution of detectors is calculated via height spectrum of waveforms with 6-us shaping time. It is necessary to know how sampling properties of a digitizer effect the energy resolution. This paper will present preliminary energy resolution results of waveforms at different sampling properties. The preliminary results show that the ENOB (effective number of bits) with 8.25-bit or better can meet the energy resolution @122keV of CDEX HPGe detectors. Based on the ADC (Analog-to-Digital Converter) quantized error theory, this paper will also make a quantitative analysis on energy resolution in CDEX HPGe detectors. It will provide guidance for ADC design in full-chain cryogenic readout electronics for HPGe detectors.