高精度分布式铀源在确定HPGe探测器有效质量与体积中的应用
Application of a high-precision distributed uranium source for determining the effective mass and volume of a HPGe detector
AI总结:
本研究利用高精度分布式铀源验证HPGe探测器的有效质量与体积,通过实验与模拟对比确认其符合制造商规格,展示了该方法在探测器质量校准中的应用价值。
摘要:
一个具有准确认证活动性的238U分布式铀源已被用于验证用于νGeN中微子实验的HPGe探测器的有效质量。该源溶解于硝酸中,为探测器晶体提供均匀的照射,允许研究其质量和体积。通过分布式源获得的实验能谱已与详细的Geant4蒙特卡洛模拟进行比较。几个伽马线的测量计数率与模拟效率一致,证实了探测器的质量和体积与制造商规格相符。结果表明,此类源可用于HPGe探测器的质量校准。
英文摘要:
A distributed uranium source with the accurately certified activity of 238U has been used to verify the effective mass of the HPGe detector intended for the νGeN neutrino experiment. The source, dissolved in nitric acid, provides homogeneous irradiation of the detector crystal allowing the study of its mass and volume. The experimental spectra obtained with the distributed source have been compared to the detailed Geant4 Monte Carlo simulations. The measured counting rates of several gamma-lines agree with the simulated efficiencies, confirming that the detector mass and volume coincide with the manufacturer specifications. The results demonstrate the applicability of such sources for mass calibration of HPGe detectors.