$γ$ - $γ$符合NAA探测痕量铀的灵敏度分析
Sensitivity analysis towards trace-uranium detection with $γ$-$γ$ coincidence NAA
摘要:
本文提出了一种改进的中子活化分析(NAA)检测痕量$^{238}$ U的方法。利用$γ$ - $γ$符合计数获得分析增强。提出了一种在大的源相关背景下评估灵敏度增益的经验方法。两个计数方案之间的检测限的比较,即,计数单伽马使用一个HPGe探测器与计数重合伽马使用两个HPGe探测器面对面放置。一个数据验证的辐射传输模型,在GEANT 4中创建一个单一的HPGe探测器,扩展到处理两个探测器的设置。在该计数方案中,与更简单的单个计数方法相比,预测圣戈班G3蓝宝石样品的$^{238}$ U检测灵敏度提高约8倍。
英文摘要:
We present an improved approach for detection of trace amounts of $^{238}$U by means of neutron activation analysis (NAA). The analysis enhancement is obtained by utilizing $γ$-$γ$ coincidence counting. An empirical method for evaluating the sensitivity gain in the presence of a large source-related background is presented. A comparison of detection limits is made between two counting schemes; namely, counting single gammas using one HPGe detector versus counting coincident gammas using two HPGe detectors placed face-to-face. A data-validated radiation transport model, created in GEANT4 for a single HPGe detector, is extended to handle the two-detector setup. In this counting scheme, a $^{238}$U-detection sensitivity enhancement of about a factor of 8 is predicted for a sample of Saint-Gobain G3 Sapphire when compared to the simpler singles counting approach.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。