通过协计中子活化分析增强蓝宝石中痕量$^{238}$U杂质的灵敏度
Enhanced sensitivity to trace $^{238}$U impurity of sapphire via coincidence neutron activation analysis
AI总结:
该研究通过γγ协计中子活化分析方法,实现了对蓝宝石中痕量$^{238}$U杂质的高灵敏度检测,确定其放射纯度。
摘要:
蓝宝石具有机械和电学性能,有利于制造辐射探测器内部组件,如时间投影室和 bolometers。然而,将其$^{232}$Th和$^{238}$U含量评估到每克皮克级别证明困难。本文报告了一项实验验证,证明γγ协计计数结合中子活化分析(NAA)可以达到皮克灵敏度。将γγ协计计数结果与先前基于单γ计数的NAA结果相结合,显示Saint Gobain蓝宝石的$^{232}$Th和$^{238}$U浓度分别为<0.26皮克和<2.3皮克,是蓝宝石放射纯度的最佳约束。
英文摘要:
Sapphire has mechanical and electrical properties that are advantageous for the construction of internal components of radiation detectors such as time projection chambers and bolometers. However, it has proved difficult to assess its $\rm ^{232}Th$ and $\rm ^{238}U$ content down to the picogram per gram level. This work reports an experimental verification of a computational study that demonstrates $γγ$ coincidence counting, coupled with neutron activation analysis (NAA), can reach ppt sensitivities. Combining results from $γγ$ coincidence counting with those of earlier single-$γ$ counting based NAA shows that a sample of Saint Gobain sapphire has $\rm ^{232}Th$ and $\rm ^{238}U$ concentrations of $<0.26$ ppt and $<2.3$ ppt, respectively; the best constraints on the radiopurity of sapphire.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。