γ谱仪在高本底活度测量中的调节
Conditioning the gamma spectrometer for activity measurement at very high background
摘要:
研究了高纯锗γ谱仪在未来反应堆燃料元件燃耗测量中的应用。HPGe探测器由Co 60源以从10 kcps到150 kcps的不同辐照率曝光,以模拟真实反应堆环境中的输入计数率。在探测器中分别放置一个Cs 137和一个Eu 152源,使其产生一定的事件率,并将Cs 137作为标记核素测量燃料元件的燃耗。结果表明,能量分辨率随辐照率的增加略有增加,高辐照水平下的透过率符合实际应用的要求。背景的影响进行了研究,在不同的参数设置中使用的特别开发的程序的背景减法。结果表明,在典型的输入辐照率和与深度燃耗相关的Cs 137强度下,本次实验中Cs 137计数率的精度始终低于2.8%,表明HPGe探测器用于未来床式反应堆燃耗测量的可行性。
英文摘要:
The application of a high purity germanium (HPGe) gamma spectrometer in determining the fuel element burnup in a future reactor is studied. The HPGe detector is exposed by a Co60 source with varying irradiation rate from 10 kcps to 150 kcps to simulate the input counting rate in real reactor environment. A Cs137 and a Eu152 source are positioned at given distances to generate certain event rate in the detector with the former being proposed as a labeling nuclide to measure the burnup of fuel element. It is shown that both the energy resolution slightly increasing with the irradiation rate and the passthrough rate at high irradiation level match the requirement of the real application. The influence of the background is studied in different parameter sets used in the particularly developed procedure of the background subtraction. It is demonstrated that with the typical input irradiation rate and Cs137 intensity relevant to deep burnup situation, the precision of the Cs137 counting rate in the current experiment is consistently below 2.8%, indicating a promising feasibility of utilizing an HPGe detector in the burnup measurement in future bed-like reactor.