The performance of the $200\times2.5\times1$ cm$^3$ plastic scintillator strip with wavelength shifting fiber read-out by two novel photodetectors called Silicon PhotoMultipliers (SiPMs) is discussed. The advantages of SiPM relative to the traditional multichannel photomultiplier are shown. Light yield and light attenuation measurements are presented. This technique can be used in muon or calorimeter systems.
Photon readout of plastic scintillators is investigated with the aim of improving the precision of neutron \b{eta} decay experiments. Neutron decay is nowadays studied with high statistics, based on up to 109 registered decay events, and leads to strongly improved limits on new physics beyond the standard model, with \b{eta}s often registered in plastic scintillators. The main systematic errors in these experiments are due to imperfect characterization of the scintillators with respect to linearity, energy resolution, and electron backscattering. We study whether these errors can be diminished…
The advent of silicon photomultipliers has enabled big advances in high energy physics instrumentation, for example by allowing the construction of extremely granular hadronic calorimeters with photon sensors integrated into small scintillator tiles. Direct coupling of the SiPM to the plastic scintillator, without use of wavelength shifting fibers, provides a fast detector response, making such devices well suited for precise timing measurements. We have constructed a setup consisting of 15 such scintillator tiles read out with fast digitizers with deep buffers to measure the time structure of…
The simulation study of the light yield and attenuation in the plastic scintillator was performed. The wavelength shifting fiber readout was embedded in the grooves machined along the entire strip surface. The scintillator strips was irradiated with a radiation source {}^{90}Sr or cosmic muons along and across the strip.
Hamamatsu R5900-00-M64 and R7600-00-M64 photomultiplier tubes will be used with wavelength--shifting optical fibres to read out scintillator strips in the MINOS near detector. We report on measurements of the gain, efficiency, linearity, crosstalk, and dark noise of 232 of these PMTs, of which 219 met MINOS requirements.
The performance of several trigger counters based on plastic scintillators with silicon photomultiplier readout is investigated with cosmic rays. Efficiency and time resolution are measured using digital waveform analysis. The obtained results are relevant for trigger subsystems of Multi-Purpose Detector (MPD) and Baryonic Matter at the Nuclotron (BM@N) at the NICA heavy-ion collider. The results show very high efficiency and good timing performance of the counters.
$Cs_2LiYCl_6:Ce^{3+}$ (CLYC) is an inorganic scintillator that has recently garnered attention for its ability to detect and discriminate between gammas and thermal neutrons. We investigate several important performance parameters of three different solid state photomultipliers (SSPMs) when reading out CLYC crystals: linearity, energy resolution, and pulse shape and discrimination ability. These performance parameters are assessed at a variety of temperatures between -20$^{\circ}$C and +50$^{\circ}$C.
Scintillators capable of detecting both neutrons and gamma-rays have generated considerable interest. In particular, the use of such scintillators with silicon photomultipliers (SiPMs) enables low-power and compact-geometry applications. Three types of Li-loaded scintillators, CLYC, CLLB, and NaIL, have been tested with a custom-designed SiPM array for temperatures between --20 and 50$^{\circ}$C. The array consists of four 6x6~mm$^2$ SiPMs arranged in a 2x2 configuration. Pulse shape discrimination is used for neutron and gamma identification. Because the pulse shape changes with temperature, …
We report on the possibility of using Silicon Photomultipliers (SiPMs) to detect the scintillation light from neutron conversion in ZnS:6LiF scintillators. The light is collected by wavelength-shifting fibers embedded into the scintillator. The difficulty of extracting neutron signals in the presence of high dark count rates of the SiPMs is addressed by applying a dedicated processing algorithm to analyze the temporal distribution of the SiPM pulses. With a single-channel prototype detection unit we demonstrate a very good neutron signal extraction at SiPM dark count rates of about 1 MHz.
We compared the performance of a SiPM array and a PMT in a laboratory setting using a single 5.08x5.08-cm cylindrical sodium iodide scintillating crystal. Photomultiplier tubes (PMTs) are the most commonly used device to monitor scintillating materials for radiation detection purposes. The systems are sometimes limited by disadvantages in the PMTs that may degrade their performance, including temperature dependence and variation with magnetic field. Instrumentation engineering must also contend with a potentially large volume relative to the active scintillator volume, fragility, and high volt…
In the evaluation of novel scintillators, it is important to ensure that the spectrum of the light emitted by the scintillator is well matched to the response of the photomultiplier. In attempting to measure this spectrum using radioactive sources, it is found that so few photons are emitted per scintillation event that conventional spectroscopic techniques cannot easily be used. A simple photon counting technique is presented, using two photomultipliers operated in coincidence, the one viewing the scintillator directly, while the other views it through a monochromator. This system allows the …
The results of a development of a scintillator counter with wavelength shifting (WLS) fibre and a multi-pixel Geiger-mode avalanche photodiode readout are presented. The photodiode has a metal-resistor-semiconductor layered structure and operates in the limited Geiger mode. The scintillator counter has been developed for the EMMA underground cosmic ray experiment.
We present light yield and uniformity measurements of square and hexagonal tiles read out with silicon photomultipliers via a Y11 wavelength-shifting fiber or directly from the side or from the center at the top face. All tiles are 3~mm thick and have an area of $\rm 9~cm^2$. The sides are wrapped with two layers of Teflon tape while top and bottom faces are covered with two layers of Tyvec paper. We further show the first light yield and uniformity measurements of ATLAS Tile Calorimeter (TileCal) tiles with MPPC readout. This study has been motivated by looking into a possible phase 3 upgrade…
Among the top priorities for high-energy astronomy in the coming decade are sensitive surveys in the hard X-ray/soft γ-ray (10-600 keV) and medium-energy γ-ray (0.2-80 MeV) bands. Historically, observations in the soft and medium energy γ-ray bands have been conducted using detectors based on inorganic scintillators read out by photo-multiplier tubes (PMTs). These observations were limited by the modest energy and time resolution of traditional scintillator materials (e.g. NaI and CsI), and by the demands on mission resources imposed by the bulky, fragile, high-voltage PMTs. Recent technologic…
The detection of 200-1000 MeV neutrons requires large amounts, $\sim$100 cm, of detector material because of the long nuclear interaction length of these particles. In the example of the NeuLAND neutron time-of-flight detector at FAIR, this is accomplished by using 3000 monolithic scintillator bars of 270$\times$5$\times$5 cm$^3$ size made of a fast plastic. Each bar is read out on the two long ends, and the needed time resolution of $σ_t$ $<$ 150 ps is reached with fast timing photomultipliers. In the present work, it is investigated whether silicon photomultiplier (SiPM) photosensors can be …
The performance of scintillator counters with embedded wavelength-shifting fibers has been measured in the Fermilab Meson Test Beam Facility using 120 GeV protons. The counters were extruded with a titanium dioxide surface coating and two channels for fibers at the Fermilab NICADD facility. Each fiber end is read out by a 2*2 mm^2 silicon photomultiplier. The signals were amplified and digitized by a custom-made front-end electronics board. Combinations of 5*2 cm^2 and 6*2 cm^2 extrusion profiles with 1.4 and 1.8 mm diameter fibers were tested. The design is intended for the cosmic-ray veto de…
Measurements are presented of the reduction of signal output due to radiation damage for plastic scintillator tiles used in the hadron endcap (HE) calorimeter of the CMS detector. The tiles were exposed to particles produced in proton-proton (pp) collisions at the CERN LHC with a center-of-mass energy of 13 TeV, corresponding to a delivered luminosity of 50 fb$^{-1}$. The measurements are based on readout channels of the HE that were instrumented with silicon photomultipliers, and are derived using data from several sources: a laser calibration system, a movable radioactive source, as well as …
Using scintillator tile technology several square meters of plastic scintillator are read out by only two photomultiplier with a time precision of about 1.5 nsec. Two examples are discussed to build a detector based on this technology to search for cosmic muons and neutrinos.
We report the first clear observation of neutron/gamma-ray pulse-shape sensitivity of a fully-instrumented 8 $\times$ 8 array of plastic scintillator segments coupled to two 5 cm $\times$ 5 cm 64-channel SiPM arrays as part of a study of the key metrics of a prototype antineutrino detector module designed for directional sensitivity. SANDD (a Segmented AntiNeutrino Directional Detector) will eventually comprise a central module of 64 elongated segments of $^{6}$Li-doped pulse-shape-sensitive scintillator rods, each with a square cross section of 5.4 mm $\times$ 5.4 mm, surrounded by larger cro…
In this work we attempt to establish the best time resolution attainable with a scintillation counter consisting of a plastic scintillator read out by a Geiger-mode Avalanche Photodiode. The measured time resolution is inversely proportional to the square root of the energy deposited in the scintillator, and scales to 18ps (sigma) at 1MeV. This result competes with the best ones reported for photomultiplier tubes.
Large-volume water-Cherenkov neutrino detectors are a light-starved environment, as each interaction produces only $\sim 50-100$ photons per MeV. As such, maximizing the light collection efficiency of the detector is vital to performance. Since Cherenkov emission is heavily weighted towards the near UV, one method to maximize overall detector light collection without increasing the number of photomultiplier tubes is to couple each tube to a wavelength-shifting plastic plate, thus shifting photon wavelengths to a regime better suited to maximize photomultiplier efficiency and potentially detect…
We present the results of the first high-altitude balloon flight test of a concept for an advanced Compton telescope making use of modern scintillator materials with silicon photomultiplier (SiPM) readouts. There is a need in the fields of high-energy astronomy and solar physics for new medium-energy gamma-ray (~0.4 - 10 MeV) detectors capable of making sensitive observations. A fast scintillator- based Compton telescope with SiPM readouts is a promising solution to this instrumentation challenge, since the fast response of the scintillators permits the rejection of background via time-of-flig…
Based on test-beam measurements, we study the response of a liquid-scintillator detector equipped with wavelength-shifting optical modules, that are proposed e.g. for the IceCube experiment and the SHiP experiment, and adiabatic light guides that are viewed either by a photomultiplier tube or by an array of silicon photomultipliers. We report on the efficiency, the time resolution and the detector response to different particle types and point out potential ways to improve the detector performance.
The Atomic Spectroscopy and Collisions Using Slow Antiprotons (ASACUSA) experiment at the Antiproton Decelerator (AD) facility of CERN constructed segmented scintillators to detect and track the charged pions which emerge from antiproton annihilations in a future superconducting radiofrequency Paul trap for antiprotons. A system of 541 cast and extruded scintillator bars were arranged in 11 detector modules which provided a spatial resolution of 17 mm. Green wavelength-shifting fibers were embedded in the scintillators, and read out by silicon photomultipliers which had a sensitive area of 1 x…
A prototype module for an International Linear Collider (ILC) detector was built, installed, and tested between 2006 and 2009 at CERN and Fermilab as part of the CALICE test beam program, in order to study the possibilities of extending energy sampling behind a hadronic calorimeter and to study the possibilities of providing muon tracking. The "tail catcher/muon tracker" (TCMT) is composed of 320 extruded scintillator strips (dimensions 1000 mm x 50 mm x 5 mm) packaged in 16 one-meter square planes interleaved between steel plates. The scintillator strips were read out with wavelength shifting…
The light yield and the time resolution of different types of 3 m long scintillating bars instrumented with wavelength shifting fibres and read out by different models of silicon photomultipliers have been measured at a test beam at the T9 area at the CERN Proton Synchrotron. The results obtained with different configurations are presented. A time resolution better than 800 ps, constant along the bar length within 20%, and a light yield of ~ 140 (70) photoelectrons are obtained for bars 3 m long, 4.5 (5) cm wide and 2 (0.7) cm thick. These results nicely match the requirements for the Muon Det…
Results are presented on R&D efforts to design and build large size veto panels, optimized for underground low background experiments, in the most efficient and economical way using commercially available components. A variety of plastic scintillators, photomultiplier tubes, wavelength shifting fibers, and light reflector combinations were tested. Results of these studies and performance of a 2.2 meter long panel are presented.
The NeuLAND (New Large-Area Neutron Detector) plastic-scintillator-based time-of-flight detector for 0.1-1.6 GeV neutrons is currently under construction at the Facility for Antiproton and Ion Research (FAIR), Darmstadt, Germany. In its final configuration, NeuLAND will consist of 3000 2.7 m $\times$ 5 cm $\times$ 5 cm big plastic scintillator bars that are read out on each end by fast timing photomultipliers. Here, data from a comprehensive study of an alternative light readout scheme using silicon photomultipliers (SiPM) are reported. For this purpose, a NeuLAND bar was instrumented on each …
Photoelectron yields of extruded scintillation counters with titanium dioxide coating and embedded wavelength shifting fibers read out by silicon photomultipliers have been measured at the Fermilab Test Beam Facility using 120\,GeV protons. The yields were measured as a function of transverse, longitudinal, and angular positions for a variety of scintillator compositions and reflective coating mixtures, fiber diameters, and photosensor sizes. Timing performance was also studied. These studies were carried out by the Cosmic Ray Veto Group of the Mu2e collaboration as part of their R\&D program.
Recent tests of a single module of the Jagiellonian Positron Emission Tomography system (J-PET) consisting of 30 cm long plastic scintillator strips have proven its applicability for the detection of annihilation quanta (0.511 MeV) with a coincidence resolving time (CRT) of 0.266 ns. The achieved resolution is almost by a factor of two better with respect to the current TOF-PET detectors and it can still be improved since, as it is shown in this article, the intrinsic limit of time resolution for the determination of time of the interaction of 0.511 MeV gamma quanta in plastic scintillators is…
We studied the temperature dependence of the light yield of the linear alkyl benzene (LAB)-based and mesitylene-based liquid scintillators. The light yield increases by 23% for both liquid scintillators when the temperature is lowered from $26\;^{\circ}$C to $-40\;^{\circ }$C, correcting for the temperature response of the photomultiplier tube. The measurements help to understand the energy response of the liquid scintillator detectors. Especially, the next generation reactor neutrino experiments for neutrino mass hierarchy, such as the Jiangmen Underground Neutrino Observatory (JUNO), require…
A high light yield calcium iodide (CaI2) scintillator is being developed for an astroparticle physics experiments. This paper reports scintillation performance of calcium iodide (CaI2) crystal. Large light emission of 2.7 times that of NaI(Tl) and an emission wavelength in good agreement with the sensitive wavelength of the photomultiplier were obtained. A study of pulse shape discrimination using alpha and gamma sources was also performed. We confirmed that CaI2 has excellent pulse shape discrimination potential with a quick analysis.
We present a novel geometry of scintillator tiles developed for fiberless coupling to silicon photomultipliers (SiPMs) for applications in highly granular calorimeters. A high degree of uniformity of the tile response over the full active area was achieved by a drilled slit at the coupling position of the photon sensor with 2 mm, 4 mm and 5.5 mm in height, width and depth. Detailed measurements of the response to penetrating electrons were performed for tiles with a lateral size of 3 x 3 cm^2 and thicknesses of 5 mm and 3 mm.
The time resolution (TR) is one of the most important characteristic of a detector. For the case of a scintillator, the collection of light also is important, it depends of the effective area of the photo-sensor (\textit{Scorer}). The Photomultipliers Tubes (PMTs), normally, have an effective area greater than the Silicon Photomultipliers (SiPMs), however, due to its voltage of operation, their size or cost, in some cases it can be difficult to work with them. In which case, it is preferable to use SiPMs. The value of TR also depends of the size and geometry of the scintillator, number of phot…
A plastic scintillator bar with dimensions 300 cm x 2.5 cm x 11 cm was exposed to a focused muon beam to study its light yield and timing characteristics as a function of position and angle of incidence. The scintillating light was read out at both ends by photomultiplier tubes whose pulse shapes were recorded by waveform digitizers. Results obtained with the WAVECATCHER and SAMPIC digitizers are analyzed and compared. A discussion of the various factors affecting the timing resolution is presented. Prospects for applications of plastic scintillator technology in large-scale particle physics d…
We describe a complete end-to-end simulation of the response of scintillator counters to charged particles. The counters were extruded with a titanium dioxide surface coating and two channels for the embedded wavelength-shifting fibers which are read out by silicon photomultipliers. The simulation includes the production and propagation of scintillation and Cerenkov photons, the response of the silicon photomultipliers, and the generation of the signal waveforms. Probability lookup tables are used to speed up the simulation of the photon propagation inside the counters. The simulation was tune…
Imaging hadronic calorimeters with scintillator readout use small scintillator tiles individually read out by silicon photomultipliers to achieve the necessary granularity needed for sophisticated reconstruction algorithms at future collider detectors. For a second generation prototype of the CALICE analog HCAL new, 3 mm thick scintillator tiles with an embedded wavelength shifting fiber and new photon sensor from CPTA are being fabricated. The availability of blue-sensitive SiPMs also allows fiberless coupling of the photon sensor to the tile, a technique requiring modified geometries to achi…
Plastic scintillator detectors have been extensively used in particle physics experiments for decades. A large-scale detector is typically arranged as an array of staggered long bars which provide a fast trigger signal and/or particle identification via time-of-flight measurements. Scintillation light is collected by photosensors coupled to both ends of every bar. In this article, we present our study on a direct replacement of commonly used vacuum photomultiplier tubes (PMTs) by arrays of large-area silicon photomultipliers (SiPMs). An SiPM array which is directly coupled to the scintillator …
Silicon photomultipliers (SiPMs) are a good alternative to photomultiplier tubes (PMTs) because their gain and quantum efficiency are comparable to PMTs. However, the largest single-chip SiPM is still less than 1~cm$^2$. In order to use SiPMs with scintillators that have reasonable sensitivity, it is necessary to use multiple SiPMs. In this work, scintillation detectors are constructed and tested with a custom 2x2 SiPM array. The layout of the SiPMs and the geometry of the scintillator were determined by performing Geant4 simulations. Cubic NaI, CsI, and CLYC with 18~mm sides have been tested.…
The results obtained in laboratory tests, using scintillator bars read by silicon photomultipliers are reported. The present approach is the first step for designing a precision tracking system to be placed inside a free magnetized volume for the charge identification of low energy crossing particles. The devised system is demonstrated able to provide a spatial resolution better than 2 mm.
An ASIC named CATIROC (Charge And Time Integrated Read Out Chip) has been developed for the next-generation neutrino experiments using a large number of photomultiplier tubes (PMTs). Each CATIROC provides the time and the charge measurements for 16 configurable input channels operating in trigger-less mode. Originally designed for the light emission in water Cherenkov detectors, we show in this paper that its use can be extended to liquid-scintillator based experiments. The $\sim$26000 3-inch PMTs of the JUNO experiment, under construction in China, is a case in point. This paper describes the…
We report on the design and commissioning of a new spectrometer for muon-spin relaxation/rotation studies installed at the Swiss Muon Source (S$μ$S) of the Paul Scherrer Institute (PSI, Switzerland). This new instrument is essentially a new design and replaces the old general-purpose surface-muon instrument (GPS) which has been for long the workhorse of the $μ$SR user facility at PSI. By making use of muon and positron detectors made of plastic scintillators read out by silicon photomultipliers (SiPMs), a time resolution of the complete instrument of about 160 ps (standard deviation) could be …
Silicon photomultipliers (SiPMs) are semiconductor-based light-sensors offering a high gain, a mechanically and optically robust design and high photon detection efficiency. Due to these characteristics, they started to replace conventional photomultiplier tubes in many applications in recent years. This paper presents an optical module based on SiPMs designed for the application in scintillators as well as lab measurements. The module hosts the SiPM bias voltage supply and three pre-amplifiers with different gain levels to exploit the full dynamic range of the SiPMs. Two SiPMs, read-out in pa…
The CALICE collaboration is presently constructing a test hadron calorimeter (HCAL) with 7620 scintillator tiles read out by novel photo-detectors - Silicon Photomultipliers (SiPMs). This prototype is the first device which uses SiPMs on a large scale. We present the design of the HCAL and report on measured properties of more than 10 thousand SiPMs. We discuss the SiPM efficiency, gain, cross-talk, and noise rate dependence on bias voltage and temperature, including the spread in these parameters. We analyze the reasons for SiPM rejection and present the results of the long term stability stu…
Compact neutron imagers using double-scatter kinematic reconstruction are being designed for localization and characterization of special nuclear material. These neutron imaging systems rely on scintillators with a rapid prompt temporal response as the detection medium. As n-p elastic scattering is the primary mechanism for light generation by fast neutron interactions in organic scintillators, proton light yield data are needed for accurate assessment of scintillator performance. The proton light yield of a series of commercial fast plastic organic scintillators---EJ-200, EJ-204, and EJ-208--…
The recent availability of blue sensitive silicon photomultipliers allows the direct readout of blue emitting plastic scintillator tiles without the use of a wavelength shifting fiber. Such directly read out tiles, without light guides, are attractive for the use in highly granular calorimeters that use large numbers of individual cells and in other applications where very compact designs are needed. However, the total signal amplitude and the uniformity of the response can be problematic in such cases. We have developed a scanning setup to investigate the response of scintillator tiles with S…
The barrel time-of-flight (TOF) detector for the PANDA experiment at FAIR in Darmstadt is planned as a scintillator tile hodoscope (SciTil) using 8000 small scintillator tiles. It will provide fast event timing for a software trigger in the otherwise trigger-less data acquisition scheme of PANDA, relative timing in a multiple track event topology as well as additional particle identification in the low momentum region. The goal is to achieve a time resolution of sigma ~ 100 ps. We have conducted measurements using organic scintillators coupled to Silicon Photomultipliers (SiPM). The results ar…
Within the CALICE collaboration, several concepts for the hadronic calorimeter of a future linear collider detector are studied. After having demonstrated the capabilities of the measurement methods in "physics prototypes", the focus now lies on improving their implementation in "engineering prototypes", that are scalable to the full linear collider detector. The Analog Hadron Calorimeter (AHCAL) concept is a sampling calorimeter of tungsten or steel absorber plates and plastic scintillator tiles read out by silicon photomultipliers (SiPMs) as active material. The front-end chips are integrate…
A large acceptance scintillator detector with wavelength shifting optical fibre readout has been designed and built to detect the decay particles of $η$-nucleus bound system (the so-called $η$-mesic nuclei), namely, protons and pions. The detector, named as ENSTAR detector, consists of 122 pieces of plastic scintillator of various shapes and sizes, which are arranged in a cylindrical geometry to provide particle identification, energy loss and coarse position information for these particles. A solid angle coverage of $\sim$95% of total 4$π$ is obtained in the present design of the detector. Mo…