Time resolution of the plastic scintillator strips with matrix photomultiplier readout for J-PET tomograph
P. Moskal, O. Rundel, D. Alfs, T. Bednarski, P. Białas, E. Czerwiński, A. Gajos, K. Giergiel, M. Gorgol, B. Jasińska, D. Kamińska, Ł. Kapłon, G. Korcyl, P. Kowa… 展开作者
P. Moskal, O. Rundel, D. Alfs, T. Bednarski, P. Białas, E. Czerwiński, A. Gajos, K. Giergiel, M. Gorgol, B. Jasińska, D. Kamińska, Ł. Kapłon, G. Korcyl, P. Kowalski, T. Kozik, W. Krzemień, E. Kubicz, Sz. Niedźwiecki, M. Pałka, L. Raczyński, Z. Rudy, N. G. Sharma, A. Słomski, M. Silarski, A. Strzelecki, A. Wieczorek, W. Wiślicki, P. Witkowski, M. Zieliński, N. Zoń
备注:
To be published in Phys. Med. Biol. (26 pages, 17 figures)
Jagiellonian正电子发射断层成像系统(J-PET)由30 cm长的塑料闪烁体条组成,最近对该系统的单个模块进行的测试已证明其适用于探测湮没量子(0.511 MeV),符合分辨时间(CRT)为0.266 ns。所实现的分辨率几乎是由一个因素的两个更好的相对于当前的TOF-PET探测器,它仍然可以改善,因为它是在这篇文章中所示的,时间分辨率的固有限制,用于确定时间的0.511 MeV的伽马量子在塑料闪烁器的相互作用是低得多。作为本文的重点,介绍了一种利用硅光电倍增器矩阵(SiPM)在闪烁体条两端记录多个光子时间戳的方法。作为模拟的结果,与SiPM的数量从4到42变化进行,它示出的定时分辨率的改善饱和与光电倍增管的数量的增加,并且在两端的2 × 5配置允许读取20个时间戳,构成了最佳的解决方案。所进行的模拟占的发射时间分布,光子传输和吸收内的闪烁体,以及量子效率和渡越时间扩散的光电传感器,并根据实验结果进行了检查。SiPM的2 x 5矩阵的应用允许在正电子发射断层成像中实现对于15 cm轴向视场(AFOV)的$\approx$ 0.170 ns和对于100 cm AFOV的$\approx$ 0.365 ns的符合分辨时间。该结果为构建具有成本效益的TOF-PET扫描仪开辟了前景,相对于当前的TOF-PET模式,该扫描仪具有显著更好的TOF分辨率和更大的AFOV。
英文摘要:
Recent tests of a single module of the Jagiellonian Positron Emission Tomography system (J-PET) consisting of 30 cm long plastic scintillator strips have proven its applicability for the detection of annihilation quanta (0.511 MeV) with a coincidence resolving time (CRT) of 0.266 ns. The achieved resolution is almost by a factor of two better with respect to the current TOF-PET detectors and it can still be improved since, as it is shown in this article, the intrinsic limit of time resolution for the determination of time of the interaction of 0.511 MeV gamma quanta in plastic scintillators is much lower. As the major point of the article, a method allowing to record timestamps of several photons, at two ends of the scintillator strip, by means of matrix of silicon photomultipliers (SiPM) is introduced. As a result of simulations, conducted with the number of SiPM varying from 4 to 42, it is shown that the improvement of timing resolution saturates with the growing number of photomultipliers, and that the 2 x 5 configuration at two ends allowing to read twenty timestamps, constitutes an optimal solution. The conducted simulations accounted for the emission time distribution, photon transport and absorption inside the scintillator, as well as quantum efficiency and transit time spread of photosensors, and were checked based on the experimental results. Application of the 2 x 5 matrix of SiPM allows for achieving the coincidence resolving time in positron emission tomography of $\approx$ 0.170 ns for 15 cm axial field-of-view (AFOV) and $\approx$ 0.365 ns for 100 cm AFOV. The results open perspectives for construction of a cost-effective TOF-PET scanner with significantly better TOF resolution and larger AFOV with respect to the current TOF-PET modalities.
Instituto de Física Corpuscular (IFIC)(粒子物理研究所);Universitat de València(瓦伦西亚大学);Donostia International Physics Center(圣塞巴斯蒂安国际物理中心);Ikerbasque (Basque Foundation for Science)(伊克尔巴斯基(巴斯克科学基金会))
Estimating the NEMA characteristics of the J-PET tomograph using the GATE package
新型的全身PET系统是由{J-PET}合作开发的。它由圆柱形轴向排列的塑料闪烁体条组成,从而实现了全身PET的成本效益结构。为了确定扫描仪原型的属性并优化其几何形状,使用GATE软件进行了高级计算机模拟。空间分辨率,灵敏度,散射分数和噪声等效计数率估计根据NEMA规范作为断层扫描仪的长度,检测层的数量,断层扫描室的直径的函数,并为各种类型的应用读出。对于直径为85 cm、条带长度为100 cm、横截面为4 mm x 20 mm的单层几何结构和具有附加层波长移位器作为读出器的硅光电倍增器,扫描器中心的空间分辨率(FWHM)等于3 mm(径向、切向)和6 mm(轴向)。对于具有相同读数、直径和闪烁体长度的类似双层几何形状,其中条带横截面为7 mm x 20 mm,在40 kBq/cc放射性浓度下达到300 kcps的NECR峰值,散射分数估计为约35%,并且中心处的灵敏度总计为14.9 cps/kBq。还测定了敏感性曲线。
A Proof-of-principle for Time-Of-Flight Positron Emission Tomography Imaging
飞行时间(TOF)是一项在全球范围内用于正电子发射断层成像(PET)成像的重要技术。全世界用于TOF-PET的基于闪烁体的成像系统非常昂贵。多间隙电阻板室(MRPC)是一种易于制造、价格低廉、具有良好位置和时间分辨率的气体探测器。它们可以用作昂贵的制冷剂的合适替代品。为了TOF-PET的唯一目的,已经制造了一对18 cm $\times$ 18 cm、5间隙、玻璃基MRPC模块。我们的主要目的是确定这些制造的MRPC的源(Na-22)的位置的变化。在本文件中,将详细介绍实验结果。
An FPGA Based energy correction method for one-to-one coupled PET detector: model and evaluation
基于硅光电倍增器(SiPMs)的PET扫描器已经被广泛用作先进的核医学成像技术,其产生局部体内生物学和生物化学的定量图像。SiPM的紧凑尺寸允许闪烁晶体和光电传感器之间的直接一对一耦合,从而产生比必须在光电倍增管(PMT)PET系统中使用的光共享方法更好的定时和能量分辨率。为了减小读出电子设备的体积,具有位置解码器的前端多路复用器是没有高度集成的专用集成电路(ASIC)的一对一系统的常见选择。然而,在这种情况下,我们不能测量每个晶体的沉积能量激发的湮灭光子,所以晶间散射(ICS)事件将导致晶体错位,然后恶化探测器的固有分辨率。此外,考虑到SiPM的增益色散和非线性输出导致的能量窗口内的事件拒绝,需要能量校正机制。然而,缺乏每个晶体的能量信息将引入大的能量校正误差的ICS事件。针对这一问题,本文提出了一种在Kintext-7现场可编程门阵列(FPGA)器件上实现的在线能量校正机制。实验室中的实验使用8 X 8分段LYSO晶体与8 X 8 SiPM(J系列,来自ON Semiconductor)阵列耦合进行,该阵列在22 Na点源激发下。测试结果表明,非ICS和ICS事件的能量都可以精确校正,能量分辨率优于12%。我们还将此方法应用于实际的临床PET扫描仪下的68 Ge线源,以验证其多通道的可靠性。
Assessment of a high-resolution candidate detector for prostate time-of-flight positron emission tomography
我们报告使用22 Na源上的MRI兼容的TOF-PET直肠内前列腺探头的探测器元件进行测量,与深度的相互作用的灵敏度。它由LYSO闪烁体晶体制成,用Lumirror包裹,通过硅光电倍增器在两端读出。通过对数据分析过程的详细描述,我们表明我们的结果指向400 ps的重合分辨时间,同时,相互作用深度分辨率为1 mm。这些吸引人的特征,以及微小的1.5 mm x 1.5 mm x 10 mm晶体尺寸,对于原型探针的实现来说非常有希望。