Estimating the NEMA characteristics of the J-PET tomograph using the GATE package

Paweł Kowalski, Wojciech Wiślicki, Roman Shopa, Lech Raczyński, Konrad Klimaszewski, Catalina Curceanu, Eryk Czerwiński, Kamil Dulski, Aleksander Gajos, Marek G… 展开作者

Paweł Kowalski, Wojciech Wiślicki, Roman Shopa, Lech Raczyński, Konrad Klimaszewski, Catalina Curceanu, Eryk Czerwiński, Kamil Dulski, Aleksander Gajos, Marek Gorgol, Neha Gupta-Sharma, Beatrix Hiesmayr, Bożena Jasińska, Łukasz Kapłon, Daria Kamińska, Grzegorz Korcyl, Tomasz Kozik, Wojciech Krzemień, Ewelina Kubicz, Muhsin Mohammed, Szymon Niedzwiecki, Marek Pałka, Monika Pawlik-Niedźwiecka, Juhi Raj, Kamil Rakoczy, Zbigniew Rudy, Sushil Sharma, Shivani Shivani, Michał Silarski, Magdalena Skurzok, Bożena Zgardzińska, Marcin Zielinski, Paweł Moskal

摘要:

新型的全身PET系统是由{J-PET}合作开发的。它由圆柱形轴向排列的塑料闪烁体条组成,从而实现了全身PET的成本效益结构。为了确定扫描仪原型的属性并优化其几何形状,使用GATE软件进行了高级计算机模拟。空间分辨率,灵敏度,散射分数和噪声等效计数率估计根据NEMA规范作为断层扫描仪的长度,检测层的数量,断层扫描室的直径的函数,并为各种类型的应用读出。对于直径为85 cm、条带长度为100 cm、横截面为4 mm x 20 mm的单层几何结构和具有附加层波长移位器作为读出器的硅光电倍增器,扫描器中心的空间分辨率(FWHM)等于3 mm(径向、切向)和6 mm(轴向)。对于具有相同读数、直径和闪烁体长度的类似双层几何形状,其中条带横截面为7 mm x 20 mm,在40 kBq/cc放射性浓度下达到300 kcps的NECR峰值,散射分数估计为约35%,并且中心处的灵敏度总计为14.9 cps/kBq。还测定了敏感性曲线。

英文摘要:

The novel whole-body PET system based on plastic scintillators is developed by the {J-PET} Collaboration. It consists of plastic scintillator strips arranged axially in the form of a cylinder, allowing the cost-effective construction of the total-body PET. In order to determine properties of the scanner prototype and optimize its geometry, advanced computer simulations using the GATE software were performed. The spatial resolution, the sensitivity, the scatter fraction and the noise equivalent count rate were estimated according to the NEMA norm as a function of the length of the tomograph, number of the detection layers, diameter of the tomographic chamber and for various types of the applied readout. For the single-layer geometry with the diameter of 85 cm, strip length of 100 cm, cross-section of 4 mm x 20 mm and silicon photomultipliers with the additional layer of wavelength shifter as the readout, the spatial resolution (FWHM) in the centre of the scanner is equal to 3 mm (radial, tangential) and 6 mm (axial). For the analogous double-layer geometry with the same readout, diameter and scintillator length, with the strip cross-section of 7 mm x 20 mm, the NECR peak of 300 kcps was reached at 40 kBq/cc activity concentration, the scatter fraction is estimated to about 35% and the sensitivity at the centre amounts to 14.9 cps/kBq. Sensitivity profiles were also determined.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑