新一代32cm轴向视野数字式Omni Legend PET-CT的NEMA NU 2-2018性能评估
NEMA NU 2-2018 performance evaluation of a new generation digital 32-cm axial field-of-view Omni Legend PET-CT
AI总结:
本研究按NEMA NU2-2018标准评估了首款商用数字锗酸铋PET-CT Omni Legend的性能,其多数指标优于同类GE扫描仪,搭配非ToF转ToF的神经网络校正后性能可媲美ToF系统,仍需临床验证优化。
摘要:
本研究对通用电气(GE)新型32cm扩展视野数字式Omni Legend PET-CT系统开展了NEMA性能评估,标志着首款商用临床数字锗酸铋技术正式推出。测试严格遵循NEMA NU2-2018标准开展,并将该系统性能与另外两款GE商用扩展视野扫描仪进行了对比,分别为数字硅酸钇镥系统(Discovery MI - 6环)和非数字锗酸铋系统(Discovery IQ)。\n针对Omni系统的评估结果显示:径向偏移1cm处的切向、径向、轴向空间分辨率半高全宽(FWHM)分别为3.76mm、3.73mm、4.25mm;系统对中心处线源的总灵敏度为44.36 cps/kBq;峰值噪声等效计数率(NECR)为501 kcps,对应活度浓度为17.8 kBq/mL;NECR峰值处的散射分数为35.48%,NEC峰值及以下的最大计数率误差为5.5%;球体对比度恢复系数范围为52%(10mm球体)至93%(37mm球体)。该系统未配备飞行时间(ToF)功能,因此未开展时间分辨率评估,PET-CT配准精度为2.4mm。\n除背景变异性(图像噪声)指标外,Omni Legend在所有NEMA测试中的性能均优于Discovery MI。飞行时间功能本身可提升信噪比,Omni系统虽无飞行时间能力,但提供了基于预训练神经网络的软件校正(训练目标为实现非ToF到ToF的转换),经该校正后,其平均性能可与ToF系统相媲美。未来需进一步验证以优化临床成像方案及该软件校正相关的超参数,并研究不同靶标尺寸下非线性校正的效果,尤其是针对真实临床扫描的场景。
英文摘要:
A NEMA performance evaluation was conducted on the new General Electric (GE) digital Omni Legend PET-CT system with 32-cm extended field-of-view. This study marks the introduction of the first-ever commercially available clinical digital bismuth germanate technology. Testing was performed in accordance with the NEMA NU2-2018 standard. A comparison was made with the performance of two other commercial GE scanners with extended fields-of-view. A digital lutetium yttrium orthosilicate system (Discovery MI - 6 ring) and a non-digital bismuth germanate system (Discovery IQ). For the Omni assessment, the tangential, radial, and axial spatial resolutions at 1 cm radial offset were measured as 3.76 mm, 3.73 mm, and 4.25 mm FWHM. The total system sensitivity to a line source at the center was 44.36 cps/kBq. The peak NECR was 501 kcps at 17.8 kBq/mL. The scatter fraction at NECR peak was 35.48%, and the maximum count-rate error at and below NEC peak was 5.5%. Sphere contrast recovery coefficients were from 52% (10 mm) to 93% (37 mm). The system does not use time of flight; thus, no assessment of timing resolution was made. The PET-CT co-registration accuracy was 2.4 mm. The performance of the Omni Legend surpassed that of the Discovery MI on all NEMA tests, except for assessments of background variability (image noise). Time of flight is associated with inherent improvements in signal-to-noise ratio. In lieu of time of flight capabilities, the Omni provides software corrections in the form of a pre-trained neural network (trained on non-ToF to ToF). With such corrections, average performance is competitive when compared to ToF systems. Further validation is required to optimize clinical imaging protocols and hyperparameters associated with such software corrections and to examine the effect of non-linear corrections with varying target size, particularly for real world, clinical scans.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。