The low cost and high resolution gas-based Multi-gap Resistive Plate Chamber (MRPC) opens a new possibility to find an efficient alternative detector for Time of Flight (TOF) based Positron Emission Tomography, where the sensitivity of the system depends largely on the time resolution of the detector. Suitable converters can be used to increase the efficiency of detection of photons from annihilation. In this work, we perform a detailed GEANT4 simulation to optimize the converter thickness thereby improving the efficiency of photon conversion. Also we have developed a Monte Carlo based simulation of MRPC response thereby obtaining the intrinsic time resolution of the detector, making it possible to simulate the final response of MRPC-based systems for PET imaging. The result of the cosmic ray test of a four-gap Bakelite-based MRPC operating in streamer mode is discussed.
The Multigap Resistive Plate Chambers (MRPCs) provide excellent timing as well as position resolutions at relatively low cost. Therefore, they can be used in medical imaging applications such as PET where precise timing is a crucial parameter of measurement. We have designed and fabricated several six-gap glass MRPCs and extensively studied their performance. In this paper, we describe the fabrication and characterization of the detector, the electronics and the data acquisition system of the setup. We present here the result of our Time Of Flight (TOF) experiment using a radioactive source Na…
The low cost and high resolution Multi-gap Resistive Plate Chamber (MRPC) opens up a new possibility to find an efficient alternative detector for the Time of Flight (TOF) based Positron Emission Tomography, where the sensitivity of the system depends largely on the time resolution of the detector. In a layered structure, suitable converters can be used to increase the photon detection efficiency. In this paper results of the cosmic ray test of a four-gap bakelite-based prototype MRPC operated in streamer mode and six-gap glass-based MRPC operated in avalanche mode are discussed.
Time-Of-Flight (TOF) is a noble technique that is used in Positron Emission Tomography (PET) imaging worldwide. The scintillator based imaging system that is being used around the world for TOF-PET is very expensive. Multi-gap Resistive Plate Chambers (MRPCs) are gaseous detectors which are easy to fabricate, inexpensive and have excellent position and timing resolution. They can be used as a suitable alternative to highly expensive scintillators. For the sole purpose of TOF-PET, a pair of 18 cm $\times$ 18 cm, 5 gap, glass-based MRPC modules have been fabricated. Our main aim was to determine…
Development of a high-granularity, high-precision timing readout electronics system for large-area MRPC detectors
Institute of Nuclear and Particle Physics, School of Physics and Astronomy, Shanghai Jiao Tong University(上海交通大学物理与天文学院核粒子物理研究所);Key Laboratory for Particle Astrophysics and Cosmology (MoE), Shanghai Key Laboratory for Particle Physics and Cosmology(教育部粒子天体物理与宇宙学重点实验室,上海粒子物理与宇宙学重点实验室);Institut de Physique des 2 Infinis de Lyon, Lyon University(里昂大学里昂双无限物理研究所);Institute of High Energy Physics, Chinese Academy of Sciences(中国科学院高能物理研究所)
本研究针对含2400个2*2 cm pad通道的大面积MRPC探测器,研发了由FEBs、时钟模块、DAQ聚合器及PETIROC2B ASIC组成的高粒度高精度定时读出系统,其单通道时间分辨率达33 ps RMS,可应用于多种定时探测器场景。
The Multi-gap Resistive Plate Chamber (MRPC) is an advanced form of Resistive Plate Chamber (RPC) detector where the gas gap is divided into sub-gaps. MRPCs are known for their good time resolution and detection efficiency for charged particles. They have found suitable applications in several high energy physics experiments like ALICE in LHC, CERN, Geneva, Switzerland, and STAR in RHIC, BNL, USA. As they have very good time resolution and are of low cost, they can be a suitable replacement for very expensive scintillators used in Positron Emission Tomography Imaging. The MRPCs that are being …
Precise quantitative delineation of tumor hypoxia is essential in radiation therapy treatment planning to improve the treatment efficacy by targeting hypoxic sub-volumes. We developed a combined imaging system of positron emission tomography (PET) and electron para-magnetic resonance imaging (EPRI) of molecular oxygen to investigate the accuracy of PET imaging in assessing tumor hypoxia. The PET/EPRI combined imaging system aims to use EPRI to precisely measure the oxygen partial pressure in tissues. This will evaluate the validity of PET hypoxic tumor imaging by (near) simultaneously acquired…
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET built from plastic scintillators. J-PET prototype consists of 192 detection modules arranged axially in three layers forming a cylindrical diagnostic chamber with the inner diameter of 85 cm and the axial field-of-view of 50 cm. An axial arrangement of long strips of plastic scintillators, their small light attenuation, superior timing properties, and relative ease of the increase of the axial field-of-view opens promising perspectives for the cost effective construction of the whole-body PET scanner, as well as construction…
We are investigating a prototype virtual pinhole positron emission tomography (VP-PET) system for small animal imaging applications. The PET detector modules were made up of 1.3 mm lutetium-yttrium oxyorthosilicate (LYSO) arrays, and the insert detectors consists of 0.6 mm pixelated Cadmium Zinc Telluride (CdZnTe). To validate the imaging experiment, we conducted a Monte Carlo simulation for the VP-PET system in Geant4 Application for Emission Tomography (GATE). For a point source of Na-22 with 0.5 mm diameter, the filtered back-projection (FBP) algorithm reconstructed PET image shows a resolu…
We report on an hybrid in-beam PET and prompt-gamma Compton imaging system aimed at quasi real-time ion-range verification in proton-therapy treatments. Proof-of-concept experiments were carried out at the radiobiology beam line of the CNA cyclotron facility using a set of two synchronous Compton imagers and different target materials. The time structure of the 18~MeV proton beam was shaped with a series of beam-on and beam-off intervals, thereby mimicking a pulsed proton beam on a long time scale. During beam-on intervals, Compton imaging was performed utilizing the high energy $γ$-rays promp…
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image based on annihilations into two photons. The main objectives of this work include: (i) estimation of the sensitivity of the total-body PET scanners for the ortho-positronium mean lifetime imaging using …
A detection system of the conventional PET tomograph is set-up to record data from e+ e- annihilation into two photons with energy of 511 keV, and it gives information on the density distribution of a radiopharmaceutical in the body of the object. In this paper we explore the possibility of performing the three gamma photons imaging based on ortho-positronium annihilation, as well as the possibility of positronium mean lifetime imaging with the J-PET tomograph constructed from plastic scintillators. For this purposes simulations of the ortho-positronium formation and its annihilation into thre…
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET device built from plastic scintillators. It is a multi-purpose detector designed for medical imaging and for studies of properties of positronium atoms in porous matter and in living organisms. In this article we report on the commissioning of the J-PET detector in view of studies of positronium decays. We present results of analysis of the positron lifetime measured in the porous polymer. The obtained results prove that J-PET is capable of performing simultaneous imaging of the density distribution of annihilation points as…
Organ-specific, targeted Field-of-View (FoV) Positron Emission Tomography (PET)/Magnetic Resonance Imaging (MRI) inserts are viable solutions for a number of imaging tasks where whole-body PET/MRI systems lack the necessary sensitivity and resolution. To meet the required PET detector performance of these systems, high count-rates and effective spatial resolutions on the order of a few mm, a novel two-axis patterned reflector foil pixelated scintillator crystal array design is developed and its proof-of-concept illustrated in-silico with the Monte Carlo radiation transport modelling toolkit Ge…
We perform a parametric study of the newly developed time-of-flight (TOF) image reconstruction algorithm, proposed for the real-time imaging in total-body Jagiellonian PET (J-PET) scanners. The asymmetric 3D filtering kernel is applied at each most likely position of electron-positron annihilation, estimated from the emissions of back-to-back $γ$-photons. The optimisation of its parameters is studied using Monte Carlo simulations of a 1-mm spherical source, NEMA IEC and XCAT phantoms inside the ideal J-PET scanner. The combination of high-pass filters which included the TOF filtered back-proje…
Single photon sensitive detectors used in high energy physics are, in some applications, required to cover areas the size of several m2, and more specifically in very strong demand with15 an ever finer imaging and timing capability for Cherenkov Ring Imaging Detector (RICH) configurations. We are evaluating the Large Area Picosecond Photo-detector (LAPPD) produced by INCOM company, as a possible candidate for future RICH detector upgrades. In this work we perform tests on the second generation device, which is capacitively coupled to a custom designed anode back plane, consisting of various pi…
Positron emission tomographs (PET) do not measure an image directly. Instead, they measure at the boundary of the field-of-view (FOV) of PET tomograph a sinogram that consists of measurements of the sums of all the counts along the lines connecting two detectors. As there is a multitude of detectors build-in typical PET tomograph structure, there are many possible detector pairs that pertain to the measurement. The problem is how to turn this measurement into an image (this is called imaging). Decisive improvement in PET image quality was reached with the introduction of iterative reconstructi…
New detector approaches in Positron Emission Tomography imaging will play an important role in reducing costs, lowering administered radiation doses, and improving overall performance. PETALO employs liquid xenon as the active scintillating medium and UV-sensitive silicon photomultipliers for scintillation readout. The scintillation time in liquid xenon is fast enough to register time-of-flight information for each detected coincidence, and sufficient scintillation is produced with low enough fluctuations to obtain good energy resolution. The present simulation study examines a full-body-sized…
The Jagiellonian positron emission tomograph (J-PET) is a multi-purpose device built out of plastic scintillators. With large acceptance and high angular resolution, it is suitable for the studies of various phenomena such as discrete symmetries in decay of positronium atom or entangled states of photons as well as medical imaging. J-PET enables the measurement of momenta together with photon polarization related observables. Large acceptance and high granularity of the J-PET detector enables measurement of ortho-positronium decays into three photons in the whole phase-space. In this paper we …
Small animal Positron Emission Tomography (PET) is dedicated to small animal imaging. Animals used in experiments, such as rats and monkeys, are often much smaller than human bodies, which requires higher position and energy precision of the PET imaging system. Besides, Flexibility, high efficiency are also the major demands of a practical PET system. These requires a high-quality analog front-end and a digital signal processing logic with high efficiency and compatibility of multiple data processing modes. The digital signal processing logic of the small animal PET system presented in this pa…