Hybrid in-beam PET- and Compton prompt-gamma imaging aimed at enhanced proton-range verification

Javier Balibrea-Correa, Jorge Lerendegui-Marco, Ion Ladarescu, Carlos Guerrero, Teresa Rodríguez-González, Maria del Carmen Jiménez-Ramos, Begoña Fernández-Mart… 展开作者

Javier Balibrea-Correa, Jorge Lerendegui-Marco, Ion Ladarescu, Carlos Guerrero, Teresa Rodríguez-González, Maria del Carmen Jiménez-Ramos, Begoña Fernández-Martínez, César Domingo-Pardo

摘要:

我们报告了一个混合在束PET和γ射线康普顿成像系统,旨在准实时离子范围验证质子治疗。在CNA回旋加速器设施的放射生物学束线上进行了概念验证实验,使用了一组两台同步康普顿成像仪和不同的靶材料。18 MeV质子束的时间结构由一系列的开束和关束间隔形成,从而在长时间尺度上模拟脉冲质子束。在射束开启间隔期间,利用从靶中发生的核反应迅速发射的高能$γ$ -射线进行康普顿成像。在射束关闭间隔的过程中,使用相同的成像仪使用活化核的$β^{+}$衰变完成原位正电子发射断层成像。所用的靶是不同材料的堆叠,也覆盖不同的质子范围和能量。这两个互补的成像技术的性能进行了系统的研究报告和实验结果的基础上解释的Monte Carlo计算。结果表明,以并行方式结合两种成像技术的可能性,其中高效率的γ射线成像与PET的高空间精度互补。由这些结果,我们建议,一个脉冲束与适当的占空比,结合在原位康普顿和PET成像可能有助于实现互补的信息和准实时范围监测的高精度。

英文摘要:

We report on an hybrid in-beam PET and prompt-gamma Compton imaging system aimed at quasi real-time ion-range verification in proton-therapy treatments. Proof-of-concept experiments were carried out at the radiobiology beam line of the CNA cyclotron facility using a set of two synchronous Compton imagers and different target materials. The time structure of the 18~MeV proton beam was shaped with a series of beam-on and beam-off intervals, thereby mimicking a pulsed proton beam on a long time scale. During beam-on intervals, Compton imaging was performed utilizing the high energy $γ$-rays promptly emitted from the nuclear reactions occurring in the targets. In the course of the beam-off intervals in-situ positron-emission tomography was accomplished with the same imagers using the $β^{+}$ decay of activated nuclei. The targets used were stacks of different materials covering also various proton ranges and energies. A systematic study on the performance of these two complementary imaging techniques is reported and the experimental results interpreted on the basis of Monte Carlo calculations. The results demonstrate the possibility to combine both imaging techniques in a concomitant way, where high-efficiency prompt-gamma imaging is complemented with the high spatial accuracy of PET. Empowered by these results we suggest that a pulsed beam with a suitable duty cycle, in conjunction with in-situ Compton- and PET-imaging may help to attain complementary information and quasi real-time range monitoring with high accuracy.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑