塑料闪烁探测器的稳健外部束校准用于体内剂量学在高剂量率近距离放射治疗中

Robust External-Beam Calibration of Plastic Scintillation Detectors for In-Vivo Dosimetry in HDR Brachytherapy

Medscint Inc.(Medscint公司)

Chahrazed Ghannoudi, Marie-Claude Lavallee, Benjamin Cote, Luc Beaulieu

AI总结:

本研究提出了一种利用6MV外部束和高光谱技术校准塑料闪烁探测器的方法,以提高近距离放射治疗中的剂量测量精度,并探讨了不确定性来源及校准有效性。

摘要:

目的:高剂量率近距离放射治疗(HDR)是一种广泛用于癌症治疗的模式。然而,它并非完全无误,存在误差和不确定性。体内剂量学(IVD)是唯一能够确认正确剂量输送的技术。本研究详细说明并验证了塑料闪烁探测器(PSD)的校准方法,绕过了近距离放射治疗校准中的剂量梯度和定位问题。方法:PRB-0057 PSD(Medscint,加拿大)被校准,1×1 mm闪烁纤维连接到20米Eska GH-4001清晰光学纤维(Mitsubishi Rayon,日本)。该纤维连接到Hyperscint-RP200研究平台以进行信号采集。在直线加速器(LINAC)上进行高光谱校准,使用6MV束流,以在近距离放射治疗测量前消除茎效应。为了验证,使用了铱-192 Flexisource(Elekta Brachy,荷兰),Sk=29447U,在电机驱动的IBA-Blue-Phantom2水箱(48×48×41cm³)中进行。在固定横向距离为1.2±0.05 cm处,以0.2 cm的步长沿源z轴以10 Hz测量剂量率。测量值与TG-43U1剂量率的相对差异(RD)被评估。与近距离放射治疗测量相关联的详细不确定性预算。结果:比较显示在1.2 cm处RD约为2.5%,对应位置不确定性小于0.15 mm。在更大的深度(最大8 cm)处,RD增加到约5%,主要由于光产量减少。不确定性取决于源-探测器距离,范围从3.81%到6.39%(k=1)在探索范围内。结论:结果证实使用6MV外部束和高光谱技术校准PSD的有效性。接近源的不确定性与位置误差一致,并由较大距离处PSD灵敏度降低主导。本研究突显了在已知不确定性面前IVD的内在限制。

英文摘要:

Purpose: HDR brachytherapy is a widely adopted modality for cancer treatment. However, it is not free from error and uncertainty. In-vivo dosimetry (IVD) is the only technique that can confirm correct dose delivery. This study details and validates a calibration method for Plastic Scintillation Detector (PSD), bypassing dose gradient and positioning issues in brachytherapy calibration. Methods: The PRB-0057 PSD (Medscint, Canada) was calibrated, 1x1 mm scintillating fiber coupled to a 20 m Eska GH-4001 clear optical fiber (Mitsubishi Rayon, Japan). The fiber is connected to the Hyperscint-RP200 research platform for signal collection. Hyperspectral calibration was performed at a LINAC with a 6 MV beam, enabling removal of stem effects before brachytherapy measurements. For validation, an Iridium-192 Flexisource (Elekta Brachy, The Netherlands), Sk=29447U, was used in a motorized IBA-Blue-Phantom2 water tank (48x48x41cm3). Dose rates were measured at 10 Hz along the source z-axis at a fixed transverse distance of 1.2+/-0.05 cm in 0.2 cm steps. Relative difference (RD) between measured and TG-43U1 dose rates was assessed. A detailed uncertainty budget was associated with brachytherapy measurements. Results: Comparison shows good agreement with RD around 2.5 $\%$ at 1.2 cm, corresponding to positional uncertainties of <0.15 mm. At greater depths up to 8 cm, RDs increase to about 5 $\%$, mainly due to reduced light yield. Uncertainties depend on the source-detector distance, ranging from 3.81 to 6.39 $\%$ (k=1) over the explored range. Conclusions: Results confirm the PSD calibration effectiveness using a 6MV external beam with hyperspectral technique. Uncertainties close to the source align with positional errors and are dominated by reduced PSD sensitivity at larger distances. The study underlined the intrinsic limitation of IVD in the face of known uncertainties.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑