Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy

Haydee M. Linares Rosales, Patricia Duguay-Drouin, Louis Archambault, Sam Beddar, Luc Beaulieu

摘要:

目的:本研究旨在优化和表征mPSD在HDR近距离放射治疗中的体内剂量学响应。研究方法:进行详尽的分析,以获得最佳的mPSD设计,最大限度地提高电离光子的相互作用产生的闪烁光收集。构建并测试了几个mPSD原型,以确定闪烁器相对于光电检测器的适当顺序,以及它们的长度作为发射的闪烁光的函数。闪烁体BCF-60、BCF-12和BCF-10构成mPSD敏感体,每个闪烁体对总光谱的贡献取决于低能范围内的辐照,为了获得最佳mPSD设计,进行了数值优化,以选择更好地匹配光发射轮廓的光学元件。优化的剂量系统用于HDR近距离治疗剂量确定。系统性能用信噪比和信号背景比来量化。结果如下:经确定,BCF-60应放置在远端位置,BCF-12应放置在中心,BCF-10应放置在相对于光电探测器的近端位置。这种配置允许通过收集光纤进行优化的光传输,避免闪烁体间激发和自吸收效应。优化的发光系统允许使用多光谱方法进行信号解卷积,在考虑切伦科夫茎效应的情况下提取每个元素的剂量。mPSD测量值与TG-43之间的差异保持在5%以下。在所有测量条件下,该系统能够正确区分产生的闪烁信号与背景信号。结论:一个mPSD的构建和优化的HDR近距离放射治疗剂量测定,使实时剂量测定,从192 Ir源高达6.5厘米。

英文摘要:

Purpose: This study aims to optimize and characterize the response of a mPSD for in vivo dosimetry in HDR brachytherapy. Methods: An exhaustive analysis was carried out in order to obtain an optimized mPSD design that maximize the scintillation light collection produced by the interaction of ionizing photons. Several mPSD prototypes were built and tested in order to determine the appropriate order of scintillators relative to the photodetector, as well as their length as a function of the scintillation light emitted. Scintillators BCF-60, BCF-12 and BCF-10 constituted the mPSD sensitive volume.Each scintillator contribution to the total spectrum was determined by irradiations in the low energy range.For the best mPSD design, a numerical optimization was done in order to select the optical components that better match the light emission profile. The optimized dosimetric system was used for HDR brachytherapy dose determination. The system performance was quantified in term of signal to noise ratio and signal to background ratio. Results: It was determined that BCF-60 should be placed at the distal position, BCF-12 in the center and BCF-10 at proximal position with respect to the photodetector.This configuration allowed for optimized light transmission through the collecting fiber, avoiding inter-scintillator excitation and self-absorption effects.The optimized luminescence system allowed for signal deconvolution using a multispectral approach, extracting the dose to each element while taking into account Cerenkov stem effect.Differences between the mPSD measurements and TG-43 remain below 5%. In all measurement conditions, the system was able to properly differentiate the produced scintillation signal from the background one. Conclusions: A mPSD was constructed and optimized for HDR brachytherapy dosimetry, enabling real time dose determination, up to 6.5cm from the 192Ir source.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑