arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2501.01341physics.med-ph
收藏

基于深度学习的长轴向视野PET扫描仪散射校正评估

Evaluation of Deep Learning-based Scatter Correction on a Long-axial Field-of-view PET scanner

Baptiste Laurent, Alexandre Bousse, Thibaut Merlin, Axel Rominger, Kuangyu Shi, Dimitris Visvikis

AI总结:

本研究评估了基于CNN U-Net的深度学习散射估计方法在长轴向视野全身PET上的性能,通过蒙特卡罗模拟训练,在体模和临床数据上均优于传统SSS方法,展现出更高的准确性、鲁棒性和病灶对比度恢复能力。

摘要:

目的:长轴向视野(LAFOV)正电子发射断层成像(PET)系统具有更高的灵敏度,由于接受角增大,探测到的响应线数量也随之增加。然而,这种增大的角度增加了多次散射的数量以及斜平面内的散射贡献。由于散射会影响重建图像的质量和定量准确性,因此必须采用比目前最先进的单次散射模拟(SSS)更精确的方法来校正这一效应,而SSS在这种扩展视野(FOV)下可能达到其性能极限。本研究是我们此前在常规PET系统上开展的基于深度学习的散射估计(DLSE)评估工作的延伸,旨在评估DLSE方法在LAFOV全身PET上的性能。方法:所提出的DLSE方法基于卷积神经网络(CNN)U-Net架构,利用发射正弦图和衰减正弦图来估计散射正弦图。该网络使用西门子Biograph Vision Quadra扫描仪模型,通过蒙特卡罗(MC)模拟XCAT体模的[18F]-FDG PET采集数据进行训练,包含多种解剖形态和剂量分布。我们首先在正弦图域和图像域中,将该方法在模拟数据上的性能与MC真值和SSS散射正弦图进行比较。随后,我们在七个[18F]-FDG和七个[18F]-PSMA临床数据集上测试了该方法,并与SSS估计结果进行比较。结果:DLSE在体模数据上表现出更高的准确性,与SSS相比对患者体型和剂量变化具有更强的鲁棒性,并且具有更好的病灶对比度恢复能力。该方法还取得了有前景的临床结果,在[18F]-FDG数据集中提高了病灶对比度,并且在未使用[18F]-PSMA进行训练的情况下,在[18F]-PSMA数据集上表现一致。

英文摘要:

Objective: Long-axial field-of-view (LAFOV) positron emission tomography (PET) systems allow higher sensitivity, with an increased number of detected lines of response induced by a larger angle of acceptance. However, this extended angle increases the number of multiple scatters and the scatter contribution within oblique planes. As scattering affects both quality and quantification of the reconstructed image, it is crucial to correct this effect with more accurate methods than the state-of-the-art single scatter simulation (SSS) that can reach its limits with such an extended field-of-view (FOV). In this work, which is an extension of our previous assessment of deep learning-based scatter estimation (DLSE) carried out on a conventional PET system, we aim to evaluate the DLSE method performance on LAFOV total-body PET. Approach: The proposed DLSE method based on a convolutional neural network (CNN) U-Net architecture uses emission and attenuation sinograms to estimate scatter sinogram. The network was trained from Monte-Carlo (MC) simulations of XCAT phantoms [18F]-FDG PET acquisitions using a Siemens Biograph Vision Quadra scanner model, with multiple morphologies and dose distributions. We firstly evaluated the method performance on simulated data in both sinogram and image domain by comparing it to the MC ground truth and SSS scatter sinograms. We then tested the method on seven [18F]-FDG and seven [18F]-PSMA clinical datasets, and compare it to SSS estimations. Results: DLSE showed superior accuracy on phantom data, greater robustness to patient size and dose variations compared to SSS, and better lesion contrast recovery. It also yielded promising clinical results, improving lesion contrasts in [18F]-FDG datasets and performing consistently with [18F]-PSMA datasets despite no training with [18F]-PSMA.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑