J-PET监测治疗质子束射程的可行性

Feasibility of the J-PET to monitor range of therapeutic proton beams

Jakub Baran, Damian Borys, Karol Brzeziński, Jan Gajewski, Michał Silarski, Neha Chug, Aurélien Coussat, Eryk Czerwiński, Meysam Dadgar, Kamil Dulski, Kavya V. … 展开作者

Jakub Baran, Damian Borys, Karol Brzeziński, Jan Gajewski, Michał Silarski, Neha Chug, Aurélien Coussat, Eryk Czerwiński, Meysam Dadgar, Kamil Dulski, Kavya V. Eliyan, Aleksander Gajos Krzysztof Kacprzak, Łukasz Kapłon, Konrad Klimaszewski, Paweł Konieczka, Renata Kopeć, Grzegorz Korcyl, Tomasz Kozik, Wojciech Krzemień, Deepak Kumar, Antony J. Lomax, Keegan McNamara, Szymon Niedźwiecki, Paweł Olko, Dominik Panek, Szymon Parzych, Elena Perez del Rio, Lech Raczyński, Moyo Simbarashe, Sushil Sharma, Shivani, Roman Y. Shopa, Tomasz Skóra, Magdalena Skurzok, Paulina Stasica, Ewa Ł. Stępień, Keyvan Tayefi, Faranak Tayefi, Damien C. Weber, Carla Winterhalter, Wojciech Wiślicki, Pawel Moskal, Antoni Rucinski

摘要:

目的:这项工作的目的是调查的可行性Jagiellonian正电子发射断层成像(J-PET)扫描仪治疗质子束范围监测。方法:使用GATE和CASToR进行PET图像重建的蒙特卡罗模拟研究,以比较6种J-PET扫描仪几何结构(3种双头和3种圆柱)。我们模拟了质子辐照的PMMA体模与一个单一的可调束(SPB)和可调布拉格峰(SOBP)的各种范围。每个扫描仪的灵敏度和精度进行了计算,并考虑到设置的成本效益,我们指出了潜在的最佳几何形状的J-PET扫描仪原型专用于质子束范围评估。主要结果:研究表明,双层圆柱形和三层双头结构是最有前途的临床应用。我们发现,扫描仪的灵敏度是10个$^{-5}$符合每一个初级质子的顺序,而SPB和SOBP照射计划的范围评估的精度被发现低于1毫米。在具有相同数量的探测器模块的扫描仪中,最好的结果被发现为三层双头几何形状。重要性:我们进行了模拟研究,证明J-PET探测器用于基于PET的质子束治疗范围监测的可行性,具有合理的灵敏度和精度,能够在临床质子治疗环境中进行临床前测试。考虑到灵敏度、精确度和成本效益,双层圆柱形和三层双头J-PET几何结构在未来的临床应用中是有希望的。需要实验测试来证实这些发现。

英文摘要:

Objective: The aim of this work is to investigate the feasibility of the Jagiellonian Positron Emission Tomography (J-PET) scanner for intra-treatment proton beam range monitoring. Approach: The Monte Carlo simulation studies with GATE and PET image reconstruction with CASToR were performed in order to compare six J-PET scanner geometries (three dual-heads and three cylindrical). We simulated proton irradiation of a PMMA phantom with a Single Pencil Beam (SPB) and Spread-Out Bragg Peak (SOBP) of various ranges. The sensitivity and precision of each scanner were calculated, and considering the setup's cost-effectiveness, we indicated potentially optimal geometries for the J-PET scanner prototype dedicated to the proton beam range assessment. Main results: The investigations indicate that the double-layer cylindrical and triple-layer double-head configurations are the most promising for clinical application. We found that the scanner sensitivity is of the order of 10$^{-5}$ coincidences per primary proton, while the precision of the range assessment for both SPB and SOBP irradiation plans was found below 1 mm. Among the scanners with the same number of detector modules, the best results are found for the triple-layer dual-head geometry. Significance: We performed simulation studies demonstrating that the feasibility of the J-PET detector for PET-based proton beam therapy range monitoring is possible with reasonable sensitivity and precision enabling its pre-clinical tests in the clinical proton therapy environment. Considering the sensitivity, precision and cost-effectiveness, the double-layer cylindrical and triple-layer dual-head J-PET geometry configurations seem promising for the future clinical application. Experimental tests are needed to confirm these findings.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑