基于J-PET技术设计的新型临床前全身PET研究:一项模拟研究

Investigation of Novel Preclinical Total Body PET Designed With J-PET Technology:A Simulation Study

M. Dadgar, S. Parzych, F. Tayefi Ardebili, J. Baran, N. Chug, C. Curceanu, E. Czerwinski, K. Dulski, K. Eliyan, A. Gajos, B. C. Hiesmayr, K. Kacprzak, L. Kaplon… 展开作者

M. Dadgar, S. Parzych, F. Tayefi Ardebili, J. Baran, N. Chug, C. Curceanu, E. Czerwinski, K. Dulski, K. Eliyan, A. Gajos, B. C. Hiesmayr, K. Kacprzak, L. Kaplon, K. Klimaszewski, P. Konieczka, G. Korcyl, T. Kozik, W. Krzemien, D. Kumar, S. Niedzwiecki, D. Panek, E. Perez del Rio, L. Raczynski, S. Sharma, Shivani, R. Y. Shopa, M. Skurzok, K. Tayefi Ardebili, S. Vandenberghe, W. Wislicki, E. Stepien, P. Moskal

AI总结:

本研究通过模拟评估了基于J-PET技术的小鼠和大鼠全身PET扫描仪,重点考察灵敏度与空间分辨率,并证明其近乎无视差设计可兼顾分辨率与灵敏度。

摘要:

人们对人用级全身正电子发射断层扫描(PET)系统的兴趣日益增长,这一兴趣同样适用于小动物研究。由于涉及药物开发和新型治疗监测的人体研究存在现有局限性,基于动物的研究成为测试和方案准备的必要步骤。在这项基于模拟的研究中,考察了两种非常规、成本效益高的小动物全身PET扫描仪(分别用于小鼠和大鼠研究),以检验它们在临床前研究中的可行性。这些扫描仪采用了雅盖隆-PET(J-PET)合作组探索的新技术进行设计。主要考察了两项关键PET特性:灵敏度和空间分辨率,以评估其性能。此外,还研究了闪烁体尺寸和飞行时间对后一参数的影响,以设计最高效的断层扫描仪。所呈现的结果表明,对于小鼠全身J-PET,可实现的系统灵敏度为2.35%,体积空间分辨率为9.46 ± 0.54 mm³;而对于大鼠全身J-PET,这两个值分别为2.6%和14.11 ± 0.80 mm³。此外,结果表明所设计的断层扫描仪几乎是无视差系统,因此解决了在提高空间分辨率和降低灵敏度之间接受准则权衡的问题。

英文摘要:

The growing interest in human-grade total body positron emission tomography (PET) systems has also application in small animal research. Due to the existing limitations in human-based studies involving drug development and novel treatment monitoring, animal-based research became a necessary step for testing and protocol preparation. In this simulation-based study two unconventional, cost-effective small animal total body PET scanners (for mouse and rat studies) have been investigated in order to inspect their feasibility for preclinical research. They were designed with the novel technology explored by the Jagiellonian-PET (J-PET) Collaboration. Two main PET characteristics: sensitivity and spatial resolution were mainly inspected to evaluate their performance. Moreover, the impact of the scintillator dimension and time-of-flight on the latter parameter was examined in order to design the most efficient tomographs. The presented results show that for mouse TB J-PET the achievable system sensitivity is equal to 2.35% and volumetric spatial resolution to 9.46 +- 0.54 mm3, while for rat TB J-PET they are equal to 2.6% and 14.11 +- 0.80 mm3, respectively. Furthermore, it was shown that the designed tomographs are almost parallax-free systems, hence, they resolve the problem of the acceptance criterion tradeoff between enhancing spatial resolution and reducing sensitivity.