利用子束强度总变差正则化提高小动物3D打印补偿器IMRT的效率

Improving the efficiency of small animal 3D printed compensator IMRT with beamlet intensity total variation regularization

Xinmin Liu, Alexander L. Van Slyke, Erik Pearson, Khayrullo Shoniyozov, Gage Redler, Rodney D. Wiersma

摘要:

目的:人们越来越关注使用现代3D打印技术在临床前规模上实施强度调制放射治疗(IMRT),这类似于临床IMRT。然而,当前的3D打印IMRT方法遭受复杂的调制模式,导致长的递送时间、过量的灯丝使用和不准确的补偿器制造。在这项工作中,我们已经开发了一个全变差正则化(TVR)的方法来解决这些问题。研究方法:TVR-IMRT是一种设计用于最小化相邻子束之间强度差异的技术,用于优化子束强度图,然后将其转换为铜掺杂PLA灯丝中相应的补偿器厚度。使用五个射束生成IMRT和TVR-IMRT计划以治疗小鼠心脏,同时保留肺组织。将单个射野剂量和复合剂量传递到胶片上,并使用伽马分析与相应的计划剂量进行比较。结果如下:与标准3D打印补偿器IMRT相比,TVR-IMRT将小射束强度和补偿器厚度的总变化减少了约50%。所消耗的补偿器材料的总质量和辐射射束开启时间减少了20- 30%,而DVH保持相当。以3%/0.3mm为标准,IMRT和TVR-IMRT的γ分析通过率分别为89.07%和95.37%。结论:TVR可以应用于小动物IMRT子束强度,以产生通量图和随后的3D打印补偿器图案,总变化较小,简化了3D打印并减少了所需的灯丝量。与传统的调强放射治疗计划相比,TVR-IMRT计划需要更少的射束开启时间,同时保持剂量符合性。

英文摘要:

Purpose: There is growing interest in the use of modern 3D printing technology to implement intensity-modulated radiation therapy (IMRT) on the preclinical scale which is analogous to clinical IMRT. However, current 3D-printed IMRT methods suffer from complex modulation patterns leading to long delivery times, excess filament usage, and inaccurate compensator fabrication. In this work, we have developed a total variation regularization (TVR) approach to address these issues. Methods: TVR-IMRT, a technique designed to minimize the intensity difference between neighboring beamlets, was used to optimize the beamlet intensity map, which was then converted to corresponding compensator thicknesses in copper-doped PLA filament. IMRT and TVR-IMRT plans using five beams were generated to treat a mouse heart while sparing lung tissue. The individual field doses and composite dose were delivered to film and compared to the corresponding planned doses using gamma analysis. Results: TVR-IMRT reduced the total variation of both the beamlet intensities and compensator thicknesses by around 50% when compared to standard 3D printed compensator IMRT. The total mass of compensator material consumed and radiation beam-on time were reduced by 20-30%, while DVHs remained comparable. Gamma analysis passing rate with 3%/0.3mm criterion was 89.07% for IMRT and 95.37% for TVR-IMRT. Conclusion: TVR can be applied to small animal IMRT beamlet intensities in order to produce fluence maps and subsequent 3D-printed compensator patterns with less total variation, simplifying 3D printing and reducing the amount of filament required. The TVR-IMRT plan required less beam-on time while maintaining the dose conformity when compared to a traditional IMRT plan.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑