3Dπ:三维正电子成像,一种使用氙掺杂液氩闪烁体的新型全身PET扫描仪

3Dπ: Three-Dimensional Positron Imaging, A Novel Total-Body PET Scanner Using Xenon-Doped Liquid Argon Scintillator

Azam Zabihi, Xinran Li, Alejandro Ramirez, Manuel D. Da Rocha Rolo, Davide Franco, Federico Gabriele, Cristiano Galbiati, Michela Lai, Daniel R. Marlow, Andrew … 展开作者

Azam Zabihi, Xinran Li, Alejandro Ramirez, Manuel D. Da Rocha Rolo, Davide Franco, Federico Gabriele, Cristiano Galbiati, Michela Lai, Daniel R. Marlow, Andrew Renshaw, Shawn Westerdale, Masayuki Wada

AI总结:

本文提出新型全身PET成像方法3Dπ,采用氙掺杂液氩闪烁体与硅光电倍增管,通过Monte Carlo模拟验证,其NECR、空间分辨率和TOF分辨率均显著提升,有望缩短扫描时间并降低辐射剂量。

摘要:

目的:本文介绍了一种名为三维正电子成像(3Dπ)的新型PET成像方法,该方法集成了全身(TB)覆盖、飞行时间(TOF)技术、超低剂量成像能力,以及受DarkSide合作项目新兴技术启发的超快读出电子学。方法:本研究基于NEMA NU 2-2018协议,使用Monte Carlo模拟评估了3Dπ的性能。该方法采用由掺氙(Xe)液氩(LAr)组成的均匀单片闪烁体,并在低温下运行硅光电倍增管(SiPM)。主要结果:观察到系统性能的显著提升,3Dπ系统实现了3.2 Mcps的噪声等效计数率(NECR),在17.3 (kBq/mL)下约为uEXPLORER峰值NECR(1.5 Mcps)的两倍。空间分辨率测量显示,在两个轴向位置上的平均FWHM为2.7 mm。系统表现出卓越的灵敏度,在视野中心放置线源时达到373 kcps/MBq。此外,3Dπ在5.3 kBq/mL下实现了151 ps的TOF分辨率,凸显了其生成低噪声高质量图像的潜力。意义:本研究强调了3Dπ在提升PET成像性能方面的潜力,有望缩短扫描时间并减少患者辐射暴露。掺Xe的LAr具有闪烁快、光产额高和成本效益好等优势。未来研究将聚焦于优化系统几何结构并进一步完善重建算法,以发挥3Dπ在临床应用中的优势。

英文摘要:

Objective: This paper introduces a novel PET imaging methodology called 3-dimensional positron imaging (3Dπ), which integrates total-body (TB) coverage, time-of-flight (TOF) technology, ultra-low dose imaging capabilities, and ultra-fast readout electronics inspired by emerging technology from the DarkSide collaboration. Approach: The study evaluates the performance of 3Dπ using Monte Carlo simulations based on NEMA NU 2-2018 protocols. The methodology employs a homogenous, monolithic scintillator composed of liquid argon (LAr) doped with xenon (Xe) with silicon photomultipliers (SiPM) operating at cryogenic temperatures. Main results: Significant enhancements in system performance are observed, with the 3Dπ system achieving a noise equivalent count rate (NECR) of 3.2 Mcps which is approximately two times higher than uEXPLORER's peak NECR (1.5 Mcps) at 17.3 (kBq/mL). Spatial resolution measurements show an average FWHM of 2.7 mm across both axial positions. The system exhibits superior sensitivity, with values reaching 373 kcps/MBq with a line source at the center of the field of view. Additionally, 3Dπ achieves a TOF resolution of 151 ps at 5.3 kBq/mL, highlighting its potential to produce high-quality images with reduced noise levels. Significance: The study underscores the potential of 3Dπ in improving PET imaging performance, offering the potential for shorter scan times and reduced radiation exposure for patients. The Xe-doped LAr offers advantages such as fast scintillation, enhanced light yield, and cost-effectiveness. Future research will focus on optimizing system geometry and further refining reconstruction algorithms to exploit the strengths of 3Dπ for clinical applications.