Jagiellonian PET扫描仪归一化方法的开发

Development of the normalization method for the Jagiellonian PET scanner

Aurélien Coussat, Wojciech Krzemien, Jakub Baran, Szymon Parzych

AI总结:

本文为雅盖隆大学研发中的Jagiellonian PET扫描仪开发了基于组件的归一化方法,以校正响应线效率差异并减少重建伪影,并通过GATE模拟与实验数据对比验证了方法的有效性。

摘要:

本工作旨在将基于组件的归一化理论应用于雅盖隆大学正在研发的Jagiellonian PET扫描仪。在任何正电子发射断层扫描采集中,沿响应线的效率会因多种物理和几何效应而变化,导致重建图像中出现严重伪影。为减轻这些效应,对每条响应线应用一个归一化系数,该系数定义为多个组件的乘积。考虑了Jagiellonian PET扫描仪几何结构的特殊性。将GATE模拟获得的结果与实验数据获得的初步结果进行了比较。

英文摘要:

This work aims at applying the theory of the component-based normalization for the Jagiellonian PET scanner, currently under development at the Jagiellonian University. In any Positron Emission Tomography acquisition, efficiency along a line-of-response can vary due to several physical and geometrical effects, leading to severe artifacts in the reconstructed image. To mitigate these effects, a normalization coefficient is applied to each line-of-response, defined as the product of several components. Specificity of the Jagiellonian PET scanner geometry is taken into account. Results obtained from GATE simulations are compared with preliminary results obtained from experimental data.