Electron FLASH platform for pre-clinical research: LINAC modification, simplification of pulse control and dosimetry

Banghao Zhou, Lixiang Guo, Weiguo Lu, Mahbubur Rahman, Rongxiao Zhang, Varghese Anto Chirayath, Yang Kyun Park, Strahinja Stojadinovic, Marvin Garza, Ken Kang-Hsin Wang

AI总结:

本文提出了一种建立基于LINAC的电子FLASH临床前研究平台的新工作流程,通过优化AFC模块实现均匀脉冲与剂量率,并验证了MC剂量引擎,为FLASH研究提供了稳健且易访问的系统方案。

摘要:

背景:FLASH放射治疗是一种在超高剂量率下向肿瘤递送治疗剂量,同时保持正常组织充分免受损伤的治疗方案。然而,全面理解其潜在机制、可能的晚期毒性及最佳分割方案对于成功的临床转化至关重要。这促使了广泛的临床前研究,使得多家研究机构启动了专门的FLASH研究项目。目的:本工作描述了建立易于访问的电子FLASH(eFLASH)平台的工作流程。该平台集成了简化的脉冲控制、优化的剂量率递送以及经过验证的Monte Carlo(MC)剂量引擎,专为FLASH临床前研究提供准确的体内剂量测定。方法:通过调整自动频率控制(AFC)模块,我们优化了LINAC脉冲形态,以实现均匀的剂量率。对6 MeV FLASH束流的MC模型进行了调试,以确保可重复的体内研究所需的准确剂量计算。结果:优化AFC模块使得能够生成均匀的脉冲形态,确保了每次脉冲的剂量一致以及整个FLASH照射过程中剂量率的均匀。MC模型与胶片测量结果高度一致。MC剂量计算表明,6 MeV FLASH足以实现小鼠全脑照射的均匀剂量分布,但对于脊髓研究可能并非最佳。结论:我们提出了一种建立基于LINAC的eFLASH研究平台的新颖工作流程,集成了优化剂量率递送技术、简化的脉冲控制系统以及经过验证的MC引擎。本工作为研究人员提供了有价值的新方法,以促进开发用于FLASH研究的稳健且易于访问的基于LINAC的系统。

英文摘要:

Background: FLASH radiotherapy is a treatment regime that delivers therapeutic dose to tumors at an ultra-high dose rate while maintaining adequate normal tissue sparing. However, a comprehensive understanding of the underlying mechanisms, potential late toxicities, and optimal fractionation schemes is important for successful clinical translation. This has necessitated extensive pre-clinical investigations, leading several research institutions to initiate dedicated FLASH research programs. Purpose: This work describes a workflow for establishing an easily accessible electron FLASH (eFLASH) platform. The platform incorporates simplified pulse control, optimized dose rate delivery, and validated Monte Carlo (MC) dose engine for accurate in vivo dosimetry dedicated to FLASH pre-clinical studies. Methods: Adjustment of the automatic frequency control (AFC) module allowed us to optimize the LINAC pulse form to achieve a uniform dose rate. A MC model for the 6 MeV FLASH beam was commissioned to ensure accurate dose calculation necessary for reproducible in vivo studies. Results: Optimizing the AFC module enabled the generation of a uniform pulse form, ensuring consistent dose per pulse and a uniform dose rate throughout FLASH irradiation. The MC model closely agreed with film measurements. MC dose calculations indicated that 6 MeV FLASH is adequate to achieve a uniform dose distribution for mouse whole brain irradiation but may not be optimal for the spinal cord study. Conclusions: We present a novel workflow for establishing a LINAC-based eFLASH research platform, incorporating techniques for optimized dose rate delivery, a simplified pulse control system, and validated MC engine. This work provides researchers with valuable new approaches to facilitate the development of robust and accessible LINAC-based system for FLASH studies.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑