用于UHDR FLASH放射治疗的二极管剂量计的表征

Characterization of a Diode Dosimeter for UHDR FLASH Radiotherapy

Mahbubur Rahman, Jakub Kozelka, Jeff Hildreth, Andreas Schonfeld, Austin M. Sloop, M. Ramish Ashraf, Petr Bruza, David J. Gladstone, Brian W. Pogue, William E. … 展开作者

Mahbubur Rahman, Jakub Kozelka, Jeff Hildreth, Andreas Schonfeld, Austin M. Sloop, M. Ramish Ashraf, Petr Bruza, David J. Gladstone, Brian W. Pogue, William E. Simon, Rongxiao Zhang

AI总结:

本研究表征了一种配备新静电计的二极管EDGE探测器,证明其在UHDR FLASH放疗中剂量线性、平均剂量率和每脉冲剂量独立性良好,并能以亚毫秒分辨率进行空间和时间剂量测定。

摘要:

目的:一种配备新设计静电计的二极管EDGE探测器已被表征用于UHDR电子束,并证明其适用于UHDR FLASH放射治疗剂量测定。方法:量化了EDGE探测器的剂量线性、平均剂量率和每脉冲剂量依赖性,并与包括W1闪烁体探测器、放射性铬胶片和电离室在内的剂量计进行了比较,这些剂量计均使用10 MeV UHDR束照射。剂量、剂量率和每脉冲剂量分别通过内部开发的基于闪烁体的反馈机制、直线加速器的重复频率和源皮距来控制。深度剂量分布和单脉冲分辨率的时间分布分别与胶片和闪烁体测量结果进行了比较。通过约5kGy的照射量化了辐射引起的响应灵敏度变化。结果:在测量范围内,EDGE探测器与胶片测量在剂量(高达70 Gy)、剂量率(接近200 Gy/s)和每脉冲剂量(高达0.63 Gy/脉冲)变化时平均分别一致在2%、5%和1%以内。探测器与W1闪烁体探测器在每脉冲剂量(高达0.78 Gy/脉冲)方面平均一致在2%以内。EDGE探测器信号与电离室(IC)测量的剂量和轫致辐射尾部的平均剂量率分别成比例,偏差在0.4%和0.2%以内。EDGE探测器测量的百分深度剂量与胶片一致在3%以内,每脉冲输出与W1闪烁体一致在-6%至+5%范围内。结论:EDGE探测器在UHDR电子束中表现出剂量线性、平均剂量率独立性和每脉冲剂量独立性。它可以在亚毫秒分辨率下在空间和时间上量化束流。

英文摘要:

Purpose: A diode EDGE Detector with a newly designed electrometer has been characterized for use in an UHDR electron beam and demonstrated appropriateness for UHDR FLASH radiotherapy dosimetry. Methods: Dose linearity, mean dose rate, and dose per pulse dependencies of the EDGE Detector were quantified and compared with dosimeters including a W1 scintillator detector, radiochromic film, and ionization chamber that were irradiated with a 10 MeV UHDR beam. The dose, dose rate and dose per pulse were controlled via an in-house developed scintillation-based feedback mechanism, repetition rate of the linear accelerator, and source-to-surface distance, respectively. Depth-dose profiles and temporal profiles at individual pulse resolution were compared to the film and scintillation measurements, respectively. The radiation-induced change in response sensitivity was quantified via irradiation of ~5kGy. Results: The EDGE Detector agreed with film measurements in the measured range with varying dose (up to 70 Gy), dose rate (nearly 200 Gy/s), and dose per pulse (up to 0.63 Gy/pulse) on average to within 2%, 5%, and 1%, respectively. The detector also agreed with W1 scintillation detector on average to within 2% for dose per pulse (up to 0.78 Gy/pulse). The EDGE Detector signal was proportional to ion chamber (IC) measured dose, and mean dose rate in the bremsstrahlung tail to within 0.4% and 0.2% respectively. The EDGE Detector measured percent depth dose agreed with film to within 3% and per pulse output agreed with W1 scintillator to within -6% to +5%. Conclusions: The EDGE Detector demonstrated dose linearity, mean dose rate independence, and dose per pulse independence for UHDR electron beams. It can quantify the beam spatially, and temporally at sub millisecond resolution.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑