一种用于超高剂量率放射治疗应用的新型时间分辨、实时闪烁剂量测定系统的表征

Characterization of a novel time-resolved, real-time scintillation dosimetry system for ultra-high dose rate radiation therapy applications

Alexander Baikalov, Daline Tho, Kevin Liu, Stefan Bartzsch, Sam Beddar, Emil Schüler

AI总结:

本文表征了一种新型实时闪烁剂量测定系统,验证其在UHDR电子束下的线性、剂量率依赖性和毫秒级单脉冲测量能力。

摘要:

背景:闪烁剂量测定在超高剂量率(UHDR)放射治疗(RT)中具有良好的应用前景,但尚无系统显示能够兼容UHDR(FLASH)-RT典型的、平均剂量率($\ar{DR}$)高于100 Gy/s且每脉冲剂量($D_p$)超过1.5 Gy的条件。本研究旨在表征一种具有适应UHDR潜力的新型闪烁体剂量测定系统。方法:在一条UHDR电子束线上对该系统进行了全面的剂量学表征。研究了该系统响应随剂量、$\ar{DR}$、$D_p$和脉冲剂量率${DR}_p$变化的函数关系,同时考察了系统剂量灵敏度(单位剂量信号)随剂量历史的变化。还评估了该系统用于时间分辨剂量读出的能力。结果:在±3%的容差范围内,该系统表现出剂量线性,并且在分别测试的1.8–1341 Gy/s和0.005–7.68 Gy范围内与$\ar{DR}$和$D_p$无关。当${DR}_p$从8.9e4 Gy/s增加到1.8e6 Gy/s时,观察到单位剂量信号降低6%。此外,连续采样光电探测器每个积分窗口所传递的剂量必须保持在0.028至11.64 Gy之间,以维持稳定的单位剂量信号响应。该系统能够准确测量以高达120 Hz传递的单个脉冲的$D_p$。在参考设置下,单位剂量信号的日间变化最高达±13%,但在每个测量日内保持一致(<±2%),并且未显示随剂量历史变化的信号损失。结论:通过每日校准和针对${DR}_p$的特定校正因子,该系统可可靠地对典型电子直线加速器产生的脉冲式常规束流和UHDR束流提供实时、毫秒级分辨的剂量测量,标志着UHDR剂量测定的重要进展。

英文摘要:

Background: Scintillation dosimetry has promising qualities for ultra-high dose rate (UHDR) radiotherapy (RT), but no system has shown compatibility with mean dose rates ($\bar{DR}$) above 100 Gy/s and doses per pulse ($D_p$) exceeding 1.5 Gy typical of UHDR (FLASH)-RT. The aim of this study was to characterize a novel scintillator dosimetry system with the potential of accommodating UHDRs. Methods: A thorough dosimetric characterization of the system was performed on an UHDR electron beamline. The system's response as a function of dose, $\bar{DR}$, $D_p$, and the pulse dose rate ${DR}_p$ was investigated, together with the system's dose sensitivity (signal per unit dose) as a function of dose history. The capabilities of the system for time-resolved dosimetric readout were also evaluated. Results: Within a tolerance of $\pm$3% the system exhibited dose linearity and was independent of $\bar{DR}$ and $D_p$ within the tested ranges of 1.8-1341 Gy/s and 0.005-7.68 Gy, respectively. A 6% reduction in the signal per unit dose was observed as ${DR}_p$ was increased from 8.9e4-1.8e6 Gy/s. Additionally, the dose delivered per integration window of the continuously sampling photodetector had to remain between 0.028 and 11.64 Gy to preserve a stable signal response per unit dose. The system accurately measured $D_p$ of individual pulses delivered at up to 120 Hz. The day-to-day variation of the signal per unit dose at a reference setup varied by up to $\pm$13% but remained consistent (<$\pm$2%) within each day of measurements and showed no signal loss as a function of dose history. Conclusions: With daily calibrations and ${DR}_p$ specific correction factors, the system reliably provides real-time, millisecond-resolved dosimetric measurements of pulsed conventional and UHDR beams from typical electron linacs, marking an important advancement in UHDR dosimetry.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑