平坦滤波自由束的可行性及剂量学的蒙特卡罗研究

Monte Carlo investigation into feasibility and dosimetry of flat Flattening Filter Free beams

Sergei Zavgorodni

摘要:

平坦化无滤波器(FFF)光束由于其非均匀性而对于较大的场尺寸是次优的。本研究的目的是探讨入射电子束的分布,将产生平坦的FFF光束不使用平坦过滤器。Monte Carlo(MC)模拟与BEAMnrc和DOSXYZnrc代码已被执行,以评估这种方法的可行性。采用Varian 21EX直线加速器头部模型(简称FF模型)模拟了6MV开放束流在水中的剂量分布。然后从FF模型中去除平坦滤波器,并使用(1)入射在靶上的6 MeV电子,(2)具有优化的电子角分布以提供尽可能平坦的剂量分布的6 MeV电子束进行MC模拟。配置(1)表示FFF波束,而配置(2)允许产生平坦FFF(F4)波束。还进行了优化,以产生一组剂量率(DR)的FF束的DR的1.25至2.4的范围内的最平坦的轮廓。计算了6MVF4束的剖面和深度剂量百分比PDD,并与FF束的结果进行了比较。计算的轮廓表明FFF光束的平坦度得到改善。事实上,直到35cm直径的圆内的场尺寸,F4光束在dmax处的平坦度比FF光束更好或相当。在离轴20 cm处,剂量从FFF的52%增加到F4射束的92%。此外,F4射束的剖面没有随深度发生显著变化,并且对于大射野,与FF射束相比,野外剂量减少约2倍。F4光束的PDD与FFF光束的PDD相似。最大建模(44 cm直径)F4射束的DR比FF射束的DR高1.25倍。结果表明,DR可以增加,同时保持光束的平坦度,但在减少字段大小的成本。

英文摘要:

Flattening filter free (FFF) beams due to their non-uniformity, are sub-optimal for larger field sizes. The purpose of this study was to investigate the incident electron beam distributions that would produce flat FFF beams without the use of flattening filter. Monte Carlo (MC) simulations with BEAMnrc and DOSXYZnrc codes have been performed to evaluate the feasibility of this approach. The dose distributions in water for open 6MV beams were simulated using Varian 21EX linac head model, which will be called flattening filter (FF) model. Flattening filter has then been removed from FF model, and MC simulations were performed using (1) 6 MeV electrons incident on the target, (2) 6 MeV electron beam with electron angular distributions optimized to provide as flat dose profiles as possible. Configuration (1) represents FFF beam while configuration (2) allowed producing a flat FFF (F4) beam. Optimizations have also been performed to produce flattest profiles for a set of dose rates (DRs) in the range from 1.25 to 2.4 of the DR of FF beam. Profiles and percentage depth doses PDDs from 6MV F4 beams have been calculated and compared to those from FF beam. Calculated profiles demonstrated improved flatness of the FFF beams. In fact, up to field sizes within the circle of 35 cm diameter the flatness of F4 beam at dmax was better or comparable to that of FF beam. At 20 cm off-axis the dose increased from 52% for FFF to 92% for F4 beam. Also, profiles of F4 beams did not change considerably with depth, and for large fields out-of-field dose was reduced by about a factor of two compared to FF beam. PDDs from F4 beams were similar to those of FFF beam. The DR for the largest modeled (44 cm diameter) F4 beam was higher than the DR from FF beam by a factor of 1.25. It was shown that the DR can be increased while maintaining beam flatness, but at the cost of reduced field size.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑