Model-Based Reconstruction for Joint Estimation of $T_{1}$, $R_{2}^{*}$ and $B_{0}$ Field Maps Using Single-Shot Inversion-Recovery Multi-Echo Radial FLASH
Xiaoqing Wang, Nick Scholand, Zhengguo Tan, Daniel Mackner, Vitali Telezki, Moritz Blumenthal, Philip Schaten, Martin Uecker
备注:
Part of this work has been presented at the ISMRM Workshop on Data Sampling and Image Reconstruction, Sedona, 2023. Submitted to Magnetic Resonance in Medicine
Purpose: To develop a model-based nonlinear reconstruction for simultaneous water-specific $T_{1}$, $R_{2}^{*}$, $B_{0}$ field and/or fat fraction (FF) mapping using single-shot inversion-recovery (IR) multi-echo radial FLASH.
Methods: The proposed model-based reconstruction jointly estimates water-specific $T_{1}$, $R_{2}^{*}$, $B_{0}$ field and/or FF maps, as well as a set of coil sensitivities directly from $k$-space obtained with a single-shot IR multi-echo radial FLASH sequence using blip gradients across echoes. Joint sparsity constraints are exploited on multiple quantitative maps to improve precision. Validations are performed on numerical and NIST phantoms and with in vivo studies of the human brain and liver at 3 T.
Results: Numerical phantom studies demonstrate the effects of fat signals in $T_{1}$ estimation and confirm good quantitative accuracy of the proposed method for all parameter maps. NIST phantom results confirm good quantitative $T_{1}$ and $R_{2}^{*}$ accuracy in comparison to Cartesian references. Apart from good quantitative accuracy and precision for multiple parameter maps, in vivo studies show improved image details utilizing the proposed joint estimation. The proposed method can achieve simultaneous water-specific $T_{1}$, $R_{2}^{*}$, $B_{0}$ field and/or FF mapping for brain (0.81 $\times$ 0.81 $\times$ 5 mm$^{3}$) and liver (1.6 $\times$ 1.6 $\times$ 6 mm$^{3}$) imaging within four seconds.
Conclusion: The proposed model-based nonlinear reconstruction, in combination with a single-shot IR multi-echo radial FLASH acquisition, enables joint estimation of accurate water-specific $T_{1}$, $R_{2}^{*}$, $B_{0}$ field and/or FF maps within four seconds. The present work is of potential value for specific clinical applications.
Rapid, High-resolution and Distortion-free $R_{2}^{*}$ Mapping of Fetal Brain using Multi-echo Radial FLASH and Model-based Reconstruction
Boston Children’s Hospital, Harvard Medical School(波士顿儿童医院,哈佛医学院);Siemens Medical Solutions(西门子医疗解决方案);University of Michigan(密歇根大学);Graz University of Technology(格拉茨工业大学);University of California Irvine(加州大学尔湾分校)
An analytical Model which Determines the Apparent T1 for Modified Look-Locker Inversion Recovery (MOLLI) -- Analysis of the Longitudinal Relaxation under the Influence of Discontinuous Balanced and Spoiled Gradient Echo Readouts
Improved liver T1rho measurement precision with a breathhold black blood single shot fast spin echo acquisition: a validation study in healthy volunteers
IQ-JEPA: A Joint-Embedding Predictive Architecture with a Hermitian Vision Transformer for Sound Speed and Attenuation Estimation from Ultrasound IQ Data
研究利用超声 IQ 数据估计声速和衰减的难题,提出 IQ-JEPA 架构,先无标签预训练编码器预测 IQ 区域潜在表示,再微调。实验表明该方法提高标签效率,自监督是主导因素,为定量超声基础模型迈出第一步。