使用反转恢复径向FLASH和无校准运动分辨基于模型重建的自由呼吸心肌T1标测

Free-Breathing Myocardial T1 Mapping using Inversion-Recovery Radial FLASH and Motion-Resolved Model-Based Reconstruction

Xiaoqing Wang, Sebastian Rosenzweig, Volkert Roeloffs, Moritz Blumenthal, Nick Scholand, Zhengguo Tan, H. Christian M. Holme, Christina Unterberg-Buchwald, Rabe… 展开作者

Xiaoqing Wang, Sebastian Rosenzweig, Volkert Roeloffs, Moritz Blumenthal, Nick Scholand, Zhengguo Tan, H. Christian M. Holme, Christina Unterberg-Buchwald, Rabea Hinkel, Martin Uecker

AI总结:

本研究提出一种自由呼吸心肌T1标测技术,结合反转恢复径向FLASH采集、改进的SSA-FARY呼吸运动估计与无校准运动分辨基于模型重建,通过解析公式校正不完整T1恢复缩短等待时间,在体模、人体和动物实验中实现了两分钟内高准确度、高精度与高重复性的动态心肌T1定量成像。

摘要:

目的:开发一种使用反转恢复(IR)径向快速小角度激发(FLASH)和无校准运动分辨基于模型重建的自由呼吸心肌T1标测技术。方法:在3T条件下,使用自由运行(自由呼吸、回顾性心脏门控)IR径向FLASH进行数据采集。首先,为减少反转之间的等待时间,推导了一个解析公式,将不完整的T1恢复考虑在内以实现准确的T1计算。其次,使用改进的奇异谱分析(SSA-FARY)技术,从对比度变化采集的k空间中心估计呼吸运动信号。第三,使用运动分辨的基于模型的重建,直接从排序后的k空间数据估计参数和线圈灵敏度图。因此,除了空间稀疏性约束外,时空总变分也可直接应用于参数图。在实验体模、十一名人类受试者和一只患有心肌梗死的幼年长白猪上进行了验证。结果:与IR自旋回波参考相比,体模结果证实了当使用新校正将等待时间从五秒减少到一秒时具有良好的T1准确性。与仅空间正则化重建相比,运动分辨的基于模型重建进一步提高了精度。除了证明使用改进的SSA-FARY可以估计可靠的呼吸运动信号外,体内研究还表明,可以在两分钟内获得具有良好精度和重复性的动态心肌T1图。结论:通过结合反转恢复径向FLASH、自门控和无校准运动分辨基于模型重建,可以在自由呼吸期间实现具有良好准确性、精度和重复性的运动分辨心肌T1标测。

英文摘要:

Purpose: To develop a free-breathing myocardial T1 mapping technique using inversion-recovery (IR) radial fast low-angle shot (FLASH) and calibrationless motion-resolved model-based reconstruction. Methods: Free-running (free-breathing, retrospective cardiac gating) IR radial FLASH is used for data acquisition at 3T. First, to reduce the waiting time between inversions, an analytical formula is derived that takes the incomplete T1 recovery into account for an accurate T1 calculation. Second, the respiratory motion signal is estimated from the k-space center of the contrast varying acquisition using an adapted singular spectrum analysis (SSA-FARY) technique. Third, a motion-resolved model-based reconstruction is used to estimate both parameter and coil sensitivity maps directly from the sorted k-space data. Thus, spatiotemporal total variation, in addition to the spatial sparsity constraints, can be directly applied to the parameter maps. Validations are performed on an experimental phantom, eleven human subjects, and a young landrace pig with myocardial infarction. Results: In comparison to an IR spin-echo reference, phantom results confirm good T1 accuracy when reducing the waiting time from five seconds to one second using the new correction. The motion-resolved model-based reconstruction further improves precision compared to the spatial regularization-only reconstruction. Aside from showing that a reliable respiratory motion signal can be estimated using modified SSA-FARY, in vivo studies demonstrate that dynamic myocardial T1 maps can be obtained within two minutes with good precision and repeatability. Conclusion: Motion-resolved myocardial T1 mapping during free-breathing with good accuracy, precision and repeatability can be achieved by combining inversion-recovery radial FLASH, self-gating and a calibrationless motion-resolved model-based reconstruction.