在3 T下进行发射B1校正的自由呼吸3D心脏T1标测
Free-breathing 3D Cardiac T1 Mapping with Transmit B1 Correction at 3T
摘要:
目的:目的:建立一种在3 T条件下进行自由呼吸3D T1标测的心脏T1标测方法。方法:建立了一种自由呼吸、ECG门控反转恢复序列,并对该序列进行了优化,以用于3 T条件下的心脏T1标测。高帧率动态图像重建稀疏(k,t)空间数据采集沿堆栈的星轨迹使用基于子空间的方法加速成像。为了提高T1映射对B1+不均匀性的鲁棒性,在T1映射中进行了T1和翻转角(FA)联合估计。在估计中利用受试者特定的数据采集时间来解释成像实验期间的自然心率变化。结果如下:仿真结果表明,T1和FA联合估计和优化的ECG门控SPGR IR采集方案可以提高T1标测的准确性和精度。体模研究表明,所提出方法和参考方法的T1图之间具有良好的一致性。以1.9 mm平面内和4.5 mm平面内空间分辨率获得健康受试者(n=6)的3D心脏T1图(40个切片),平均成像时间为14.2 ± 1.6 min(心率:64.2 ± 7.1 bpm),显示心肌T1值与MOLLI获得的值相当。所提出的方法产生的B1+地图与空间平滑的变化,显示21-32%和11-15%的变化,跨室间隔外侧和冠状动脉前区的心肌在左心室。结论:所提出的方法允许在实际成像时间内进行全心脏的自由呼吸3D T1标测和发射B1校正。
英文摘要:
Purpose: To develop a cardiac T1 mapping method for free-breathing 3D T1 mapping of the whole heart at 3T with transmit B1 (B1+) correction Methods: A free-breathing, ECG-gated inversion recovery sequence with spoiled gradient-echo readout was developed and optimized for cardiac T1 mapping at 3T. High-frame rate dynamic images were reconstructed from sparse (k,t)-space data acquired along a stack-of-stars trajectory using a subspace-based method for accelerated imaging. Joint T1 and flip-angle (FA) estimation was performed in T1 mapping to improve its robustness to B1+ inhomogeneity. Subject-specific timing of data acquisition was utilized in the estimation to account for natural heart-rate variations during the imaging experiment. Results: Simulations showed that accuracy and precision of T1 mapping can be improved with joint T1 and FA estimation and optimized ECG-gated SPGR-based IR acquisition scheme. The phantom study showed good agreement between the T1 maps from the proposed method and the reference method. 3D cardiac T1 maps (40 slices) were obtained at a 1.9 mm in-plane and 4.5 mm through-plane spatial resolution from healthy subjects (n=6) with an average imaging time of 14.2 +- 1.6 min (heartbeat rate: 64.2 +- 7.1 bpm), showing myocardial T1 values comparable to those obtained from MOLLI. The proposed method generated B1+ maps with spatially smooth variation showing 21-32% and 11-15% variations across the septal-lateral and inferior-anterior regions of the myocardium in the left ventricle. Conclusion: The proposed method allows free-breathing 3D T1 mapping of the whole-heart with transmit B1 correction in a practical imaging time.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。