Real-time cardiac cine MRI -- A comparison of a diffusion probabilistic model with alternative state-of-the-art image reconstruction techniques for undersampled spiral acquisitions
Oliver Schad, Julius Frederik Heidenreich, Nils-Christian Petri, Jonas Kleineisel, Simon Sauer, Thorsten Bley, Peter Nordbeck, Bernhard Petritsch, Tobias Wech
ECG-gated cine imaging in breath-hold enables high-quality diagnostics in most patients, arrhythmia and inability to hold breath, however, can severely corrupt outcomes. Real-time cardiac MRI in free-breathing leverages robust and faster investigations regardless of these confounding factors. With the need for sufficient acceleration, adequate reconstruction methods, which transfer data into high quality images, are required. Undersampled spiral real-time acquisitions in free-breathing were conducted in a study with 16 healthy volunteers and 5 patients. Image reconstructions were performed using a novel score-based diffusion model, as well as a variational network and different compressed sensing approaches. The techniques were compared by means of an expert reader study, by calculating scalar metrics and difference images with respect to a segmented reference, and by a Bland-Altman analysis of cardiac functional parameters. In participants with irregular RR-cycles, spiral real-time acquisitions showed superior image quality with respect to the clinical reference standard. Reconstructions using the diffusion model, the variational network and l1-wavelets offered an overall comparable image quality, however sharpness was slightly increased by the diffusion approach. While slightly larger ejection fractions for the real-time acquisitions were exhibited with a bias of 1.6% for healthy subjects, differences in the data acquisition procedure resulted in uncertainties of 7.4%. The proposed real-time technique enables free-breathing acquisitions of spatio-temporal images with high-quality, covering the entire heart in less than one minute. Evaluation of the ejection fractions using the segmented reference can be significantly corrupted due to arrhythmias and averaging effects. Prolonged inference times of the diffusion model represent the main obstacle to overcome for clinical translation.
Cardiac cine magnetic resonance imaging (MRI) can be considered the optimal criterion for measuring cardiac function. This imaging technique can provide us with detailed information about cardiac structure, tissue composition and even blood flow. This work considers the application of the higher order dynamic mode decomposition (HODMD) method to a set of MR images of a heart, with the ultimate goal of identifying the main patterns and frequencies driving the heart dynamics. A novel algorithm based on singular value decomposition combined with HODMD is introduced, providing a three-dimensional …
Purpose: To develop a method for banding-free bSSFP cardiac cine with substantially reduced flow artifacts. Methods: A projection-reconstruction (PR) trajectory is proposed for a frequency-modulated cine sequence, facilitating reconstruction of three phase cycles and a field-map time series from a short, breath-held scan. Data is also acquired during the gradient rewinders to enable generation of field maps using BMART, B$_0$ mapping using rewinding trajectories, where the rewind data forms the second TE image for calculating the field map. A field-map-based combination method is developed whi…
Rensselaer Polytechnic Institute(伦斯勒理工学院);Cedars-Sinai Medical Center(西达赛奈医疗中心);University of California, Los Angeles(加州大学洛杉矶分校)
Cine cardiac magnetic resonance imaging (MRI) is widely used for diagnosis of cardiac diseases thanks to its ability to present cardiovascular features in excellent contrast. As compared to computed tomography (CT), MRI, however, requires a long scan time, which inevitably induces motion artifacts and causes patients' discomfort. Thus, there has been a strong clinical motivation to develop techniques to reduce both the scan time and motion artifacts. Given its successful applications in other medical imaging tasks such as MRI super-resolution and CT metal artifact reduction, deep learning is a…
WarpPINN-fibers: improved cardiac strain estimation from cine-MR with physics-informed neural networks
Millennium Institute for Intelligent Healthcare Engineering, iHEALTH;Department of Mechanical;Metallurgical Engineering, School of Engineering, Pontificia Universidad Cat\'olica de Chile, Santiago, Chile;Institute for Biological;Medical Engineering, Schools of Engineering, Medicine
Purpose: To develop a fully automatic method for extraction and directionality determination of respiratory signal in free-breathing, real-time (RT) cardiac MRI. Methods: The respiratory signal is extracted by a principal component analysis method from RT cine images. Then, a two-step procedure is used to determine the directionality (sign) of the respiratory signal. First, the signal polarity of all slices is made consistent with a reference slice. Second, a global sign correction is performed by maximizing the correlation of the respiratory signal with the zeroth-moment center curve. The pro…
Department Artificial Intelligence in Biomedical Engineering, Friedrich-Alexander-Universität Erlangen-Nürnberg(生物医学工程人工智能系,埃尔朗根-纽伦堡弗里德里希-亚历山大大学);Department of Biomedical Engineering, The Ohio State University(生物医学工程系,俄亥俄州立大学);Research & Clinical Translation, Magnetic Resonance, Siemens Healthineers AG(研究与临床转化,磁共振,西门子医疗有限公司);Division of Cardiovascular Medicine, The Ohio State University(心血管医学系,俄亥俄州立大学)
Motion free reconstruction of compressively sampled cardiac perfusion MR images is a challenging problem. It is due to the aliasing artifacts and the rapid contrast changes in the reconstructed perfusion images. In addition to the reconstruction limitations, many registration algorithms under perform in the presence of the rapid intensity changes. In this paper, we propose a novel motion correction method that reconstructs the motion free image series from the undersampled cardiac perfusion MR data. The motion correction method uses the novel robust principal component analysis based reconstru…
Heart failure is a leading cause of mortality in the Western world. The mouse is a widely used model for a number of diseases, induced by genetic modification or surgical intervention. When performing experiments in mice, in vivo magnetic resonance imaging (MRI) can be used to evaluate heart anatomy and function at multiple levels. Volumetric measurements of ventricle size at each phase of the heart cycle (obtained with cine MRI) can be used as a sensitive measure for heart failure. Cine MRI is applied here to genetic models of heart disease, including stress tests with pharmacological manipul…
Imperial College London(帝国理工学院);Royal Brompton Hospital(皇家布朗普顿医院)
Segmentation of cardiac fibrosis and scar are essential for clinical diagnosis and can provide invaluable guidance for the treatment of cardiac diseases. Late Gadolinium enhancement (LGE) cardiovascular magnetic resonance (CMR) has been successful for its efficacy in guiding the clinical diagnosis and treatment reliably. For LGE CMR, many methods have demonstrated success in accurately segmenting scarring regions. Co-registration with other non-contrast-agent (non-CA) modalities, balanced steady-state free precession (bSSFP) and cine magnetic resonance imaging (MRI) for example, can further en…
Deep Learning Assisted Outer Volume Removal for Highly-Accelerated Real-Time Dynamic MRI
Department of Electrical and Computer Engineering, University of Minnesota(电气与计算机工程系,明尼苏达大学);Center for Magnetic Resonance Research, University of Minnesota(磁共振研究中心,明尼苏达大学);Department of Imaging Physics, Delft University of Technology(成像物理系,代尔夫特理工大学);HollandPTC, Delft, the Netherlands(HollandPTC,代尔夫特,荷兰)
GE Global Research Center(GE全球研究中心);Technische Universität München(慕尼黑工业大学);INSERM(法国国家健康与医学研究院);Université de Lorraine(洛林大学);GE Healthcare(GE医疗)
To develop an efficient motion-compensated reconstruction technique for free-breathing cardiac magnetic resonance imaging (MRI) that allows high-quality images to be reconstructed from multiple undersampled single-shot acquisitions. The proposed method is a joint image reconstruction and motion correction method consisting of several steps, including a non-rigid motion extraction and a motion-compensated reconstruction. The reconstruction includes a denoising with the Beltrami regularization, which offers an ideal compromise between feature preservation and staircasing reduction. Results were …