快速、高分辨率且无失真的胎儿脑$R_{2}^{*}$映射:使用多回波径向FLASH和基于模型的重建

Rapid, High-resolution and Distortion-free $R_{2}^{*}$ Mapping of Fetal Brain using Multi-echo Radial FLASH and Model-based Reconstruction

Boston Children’s Hospital, Harvard Medical School(波士顿儿童医院,哈佛医学院);Siemens Medical Solutions(西门子医疗解决方案);University of Michigan(密歇根大学);Graz University of Technology(格拉茨工业大学);University of California Irvine(加州大学尔湾分校)

Xiaoqing Wang, Hongli Fan, Zhengguo Tan, Serge Vasylechko, Edward Yang, Ryne Didier, Onur Afacan, Martin Uecker, Simon K. Warfield, Ali Gholipour

AI总结:

本文提出一种快速、高分辨率且无失真的技术,用于在3T下同时分离水-脂肪、$R_{2}^{*}$和$B_{0}$映射。通过多回波径向FLASH序列和基于模型的重建方法,实现高精度的胎儿脑$R_{2}^{*}$映射,具有更高的分辨率和更低的失真。

摘要:

目的:开发一种快速、高分辨率且无失真的技术,用于在3T下同时分离水-脂肪、$R_{2}^{*}$和$B_{0}$映射。方法:采用2D多回波径向FLASH序列结合blip梯度进行数据采集,适用于孕妇自由呼吸期间。开发了一种无需校准的基于模型的重建方法,结合稀疏约束,直接从k空间估计水、脂肪、$R_{2}^{*}$和$B_{0}$场映射。该方法通过数值和NIST假体以及九例26至36周胎儿的数据验证和比较参考方法。结果:数值和实验假体研究均证实了良好的准确性和精度。在胎儿研究中,基于模型的重建方法得到的$R_{2}^{*}$值与使用图割(内类相关系数[ICC] = 0.9601)的并行成像压缩感知(PICS)技术结果接近,同时提供更详细的图像。重复扫描证实了良好的可重复性(ICC = 0.9213)。与多回波EPI相比,所提出的径向技术产生了更高分辨率(1.1×1.1×3 mm³ vs. 2-3×2-3×3 mm³)的$R_{2}^{*}$图,并减少了失真。尽管在运动、分辨率和失真方面存在差异,两种获取策略之间的$R_{2}^{*}$值仍具有可比性(ICC = 0.8049)。此外,所提出的方法能够合成高分辨率且无失真的$R_{2}^{*}$加权图像。结论:本研究展示了使用多回波径向FLASH结合无需校准的基于模型的重建方法,在3T下实现胎儿脑$R_{2}^{*}$映射的可行性,实现了2秒每切片内约1.1×1.1×3 mm³的名义分辨率。

英文摘要:

Purpose: To develop a rapid, high-resolution and distortion-free technique for simultaneous water-fat separation, $R_{2}^{*}$ and $B_{0}$ mapping of the fetal brain at 3T. Methods: A 2D multi-echo radial FLASH sequence with blip gradients is adapted for data acquisition during maternal free breathing. A calibrationless model-based reconstruction with sparsity constraints is developed to jointly estimate water, fat, $R_{2}^{*}$ and $B_{0}$ field maps directly from k-space. This approach was validated and compared to reference methods using numerical and NIST phantoms and data from nine fetuses between 26 and 36 weeks of gestation age. Results: Both numerical and experimental phantom studies confirm good accuracy and precision. In fetal studies, model-based reconstruction yields quantitative $R_{2}^{*}$ values in close agreement with those from a parallel imaging compressed sensing (PICS) technique using Graph Cut (intra-class correlation coefficient [ICC] = 0.9601), while providing enhanced image detail. Repeated scans confirm good reproducibility (ICC = 0.9213). Compared to multi-echo EPI, the proposed radial technique produces higher-resolution (1.1 $\times$ 1.1 $\times$ 3 mm$^{3}$ vs. 2-3 $\times$ 2-3 $\times$ 3 mm$^{3}$) $R_{2}^{*}$ maps with reduced distortion. Despite of differences in motion, resolution and distortion, $R_{2}^{*}$ values are comparable between the two acquisition strategies (ICC = 0.8049). Additionally, the proposed approach enables synthesis of high-resolution and distortion-free $R_{2}^{*}$-weighted images. Conclusion: This study demonstrates the feasibility of using multi-echo radial FLASH combined with calibrationless model-based reconstruction for motion-robust, distortion-free $R_{2}^{*}$ mapping of the fetal brain at 3T, achieving a nominal resolution of $1.1 \times 1.1 \times 3$ mm$^{3}$ within 2 seconds per slice.